Environmental Control
Temperature Control
7-16
Using the LI-6400
7
Temperature Control
Temperature control is achieved using dual Peltier devices on the sides of the
IRGA / sensor head. These devices heat or cool the air that is circulated
through the leaf chamber.
When
f4
(level 2) is pressed in New Measurements Mode, the following con-
trol options are presented:
Constant Block Temperature
Block temperature is measured in the wall of the IRGA, and is primarily used
for the feedback control on the coolers. This control loop is fairly straightfor-
ward, and should be able to hold temperatures at targets that are within 7 de-
grees of ambient (larger if warming, since warming is more efficient than
cooling).
Constant Leaf Temperature
The constant leaf temperature option is not a tight control loop, for two rea-
sons: 1) the control of leaf temperature is indirect, via air temperature, and 2)
there are factors beyond the reach of the controller that affect leaf tempera-
ture, including leaf transpiration rate and incident radiation.
Table 7-3.
Temperature control options
Option
Description
N) None
Turns off the chamber coolers.
B) Block Temp
Maintains a constant block temperature. Good default
option.
L) Leaf Temp
Maintains a constant leaf temperature.
N) None (Cooler Off)
B) Block temp 20 C
L) Leaf temp 20 C
Temp Control Options & Targets
Figure 7-14. The temperature control screen provides for block or leaf tem-
perature control
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...