Calibration Issues
6400-01 CO2 Mixer
Using the LI-6400
18-23
4
Automatic mixer calibration
The program loops through 8 set points, from 5000 mV down to 0 mV, and
records the CO
2
concentration for each (Figure 18-14).
Generally, it should take about 20 to 30 seconds to do each set point. The cri-
teria is that
Rangeµml
must be < 1, and the Status must be OK. (The last set
point, 0 mV, ignores the Status). Pressing
f1
(
earlyOK
) causes the point to be
accepted, and it moves to the next set point.
5
View the calibration curve
Once the calibration is done, the data is printed and
Plot this (Y/N)
is asked. Press
Y
. You should see something like Figure 18-15. (You can also
see this plot at any later time, by selecting
"_CO2 Mixer - Plot Curve"
from
the Calib Menu.)
If the user accepts the calibration, it is stored in
/User/Configs/Mixer
, along with
ÒCONFIG.JoeÓ
CO2R_µml Status Flow_µml Rangeµml
2330.1 OK 560.1 18.55
1st set pt: 5000mV Time: 4
abort
Please Stand By...
Setting the mixer at 8 different set
points, and measuring CO2R at each.
earlyOK
Figure 18-14. While doing the automatic calibration, the display will show CO2 mixer set point value
and resulting CO2 concentration.
Set point #
and value
Stability is reached
when the range is <1,
and the Status is OK
Figure 18-15. Typical plot of a CO2 mixers calibration curve. The
range of CO
2
concentrations is typically 40 to 2200 µmol mol
-1
.
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...