28
Lubrication Schedule
LFA recommends the following PharmaCap
™
Capsule Filler parts to be lubricated according to
the following frequency:
Part
Location
Image
Frequency
Type of Lubricant
Tamping Plate
Handle
The handle threads
on top of the
tamping plate that
locks it into place
After cleaning
Mineral oil
Eccentric Pin
Handle
Handle on the side
that is pulled to
separate capsule
caps and bodies in
the PharmaCap
™
After cleaning
Mineral oil
Large Lever
The point at which
the side large
lever hinges on the
PharmaCap
™
Every 100 hours/once
per week
Mineral oil
Eccentric Pin
Pin near the large
lever that goes
through the top
section
Every 2-3 months or
after cleaning
NLGI Grade 2
Cogs
Underneath the
PharmaCap
™
Every 2-3 months or
after cleaning
NLGI Grade 1