2

Installation / Connection

Caution

Connect and disconnect the cables only while
the pump is at standstill and with the mains
power switched off. Plugging cables in or out
with the mains power present or while the pump
is still running can severely damage the fre-
quency converter.

The TURBO.POWER 300 can be set up on the benchtop. After
having unscrewed the feet, the unit may be installed in electri-
cal cabinets, for example. It can be mounted using either the
grooves and bores at the short sides in the bottom panel or
with the aid of an adaptor the unit can be affixed to hat rails.
Attach the hat rail adaptor to the bottom panel using screws.
You may order the hat rail adaptor from Leybold.

Caution

When mounting the TURBO.POWER 300 

do

not make use of the holes for the feet.

The TURBO.POWER 300 is equipped with a fan. The cooling
air enters from the top and exits the housing at the side. Main-
tain a clearance of at least 3 cm at the top and at the sides.

Connect and affix the connecting cable between frequency
converter and TURBO. POWER 300.

The TURBO.POWER 300 is not equipped with a mains switch.
As soon as the mains cord has been connected, the unit will
power up.

3

Operation

Warning

The turbomolecular pump must only be operated
if installed in compliance with the information
provided in the Operating Instructions for the tur-
bomolecular pump and the frequency converter.

Connect the mains cord. The mains power circuit must be of
the earthed type.

Switch the turbomolecular pump on and off (see Operating
Instructions for the frequency converter and the turbomolecular
pump).

4

Troubleshooting

10

GA 05.280/1 - 01/2002

Symptom

After having
applied the mains
power the fre-
quency converter
does not receive
any power.

Likely cause

Remedy

AC fuse blown. 

Replace fuse. 

Cables not properly  Check the
cables.
connected.

Содержание TURBO.POWER 300

Страница 1: ...r Turbo Molekularpumpen TURBOVAC Power Supply for TURBOVAC Turbomolecular Pumps Kat Nr Part No 800100V0002 Operating Instructions Gebrauchsanleitung GA 05 280 1 Vakuum L sungen Applikations Unterst t...

Страница 2: ...300H The mains input is protected by two fuses 1 1 Ordering Information TURBO POWER 300 800100V0002 Supplies the TURBO DRIVE S with 24 V DC Plug play Bench top unit or for cabinet mounting 24V DC powe...

Страница 3: ...8 GA 05 280 1 01 2002 246 266 116 4 5 70 116 53 4 Fig 1 Dimensional drawing for TURBO POWER 300 dimensions in mm...

Страница 4: ...g System of protection EN 60529 IP30 Ambient temperature during operation 0 40 C in storage 20 70 C Humidity class F non condensing Cooling built in fan Resistance to interference EN 61000 6 2 Industr...

Страница 5: ...ed with a fan The cooling air enters from the top and exits the housing at the side Main tain a clearance of at least 3 cm at the top and at the sides Connect and affix the connecting cable between fr...

Страница 6: ...esi gned to comply with the applicable EC directive Any changes on the power supply not agreed upon with us will void this declaration Equipment TURBO POWER 300 Power Supply for Turbomolecular Pumps P...

Страница 7: ...asse 498 Bayenthal D 50968 K ln Tel 0221 347 0 Fax 0221 347 1250 http www leyboldvac de e mail documentation leyboldvac de AMCL_04411_2002 PG 01 02 Printed in Germany on chlorine free bleached paper T...

Отзывы: