Replace the gauge tube
If the gauge tube has become severely contaminated so that atmosphere or vacuum adjustment cannot be achieved,
or if the filament is broken, then you can fit a replacement tube to the gauge.
Refer to the below figure and follow this procedure to replace the gauge tube.
Unplug the electrical cable, vent the vacuum system to atmospheric pressure and remove the gauge from the
vacuum system.
Pull the retaining clip from side of gauge.
Pull the tube from the electronics housing.
Fit the replacement tube into electronics housing, noting the correct alignment.
Refit the retaining clip.
Whenever a new tube is fitted it is n
ecessary to adjust the gauge to match the new tube. Refer to ‘Adjustment for
new tube’ above.
Replacement of gauge tube
1
Electronic housing
2
Retaining clip
3
Gauge tube