Installation
25
300407956_002_C1 - 01/2017 - © Leybold
3 Installation
3.1 Ambient Requirements and Placement
The pump is designed for operation in buildings with a good ventilation. It
must not be installed in sites which are highly-polluted, very humid, full of
corrosive gases, and metal dust. The pump must not be exposed to direct
sunlight or water.
For maintenance purposes the clearance on top and around the pump must
be at least 900 mm.
The ambient temperature for the pump should not exceed 45 °C. The install-
ing site should not be near boilers or any other equipment that would be
heat-radiating.
Ensure a good ventilation and heat dissipation with the pump installed
indoors; therefore the location for the vacuum pump should be considered
cautiously.
For a convenient installation, four mobile support units, each consisting of a
castor and a height-adjustment foot, are attached at the frame of the pump.
Before moving the pump, make sure to check the four adjustment feet are
already at their highest positions.
The pump should be installed on a floor which can stand the weight of the
pump. A shock-absorbing pad should be used between the frame of the
pump and the floor for a better ground holding and less impact to avoid
vibration.
Place the pump system on a flat and level surface. For proper securing
against rolling away, use all four adjustable feet for aligning thus taking away
the load from the castors.
Adjust the adjustment feet on the pump frame until the moveable wheels are
3-5 mm off the ground; then tighten the M16 nuts to fix the pump.
Remove covers and blank flanges on the pump only before fitting the pump
so that the assembly work can be performed under the cleanest conditions.
The pumps are supplied filled with PFPE (LVO 410) or synthetic oil (LVO 210).
Nothing will have to be refilled. Check the oil levels through all oil level glass-
es. If one of the oil levels is found to be incorrect, please contact us.
The oil levels of both, the gear box and the low pressure side, should be
between 1/2 - 2/3 of the oil level glass. If the level is below 1/2, lubricant
should be added. See the Maintenance section for further information.
Fig. 3.1 Adjusting Castors