48
GA130001441_002_C0
– 09/2016
8.3.15
SSC (Set Sensor Control
Setting the type of sensor control for PENNINGVAC.
S:
SSC[a, b, c, x.xxxxE±xx, x.xxxxE±xx]<CR>
E:
OK<CR>
Parameter
Description
a
Channel number
3 = Channel 3 (PM)
b
Sensor switch-on type channel
0 = Manual
1 = External
2 = n.a.
3 = Through channel 1 (TM1)
4 = Through channel 2 (TM2)
c
Sensor switch-off type channel
0 = Manual
1 = External
2 = Self-monitoring
3 = Through channel 1 (TM1)
4 = Through channel 2 (TM2)
x.xxxxE±xx
Switch-on value in the current unit of measurement
x.xxxxE±xx
Switch-off value in the current unit of measurement
8.3.16
RSP (Read Set Point)
Reading the switching thresholds for a channel.
S:
RSP[a]<CR>
E:
b[,][TAB]c[,][TAB]d[,][TAB]e<CR>
Parameter
Description
a
Channel number
1 = Channel 1 (TM1)
2 = Channel 2 (TM2)
3 = Channel 3 (PM)
b
SP1 low (pressure value of the switching threshold
in the current unit of measurement in the form of
x.xxxxE±xx)
c
SP1 high (pressure value of the switching threshold
in the current unit of measurement in the form of
x.xxxxE±xx)
d
SP2 low (pressure value of the switching threshold
in the current unit of measurement in the form of
x.xxxxE±xx)
e
SP2 high (pressure value of the switching threshold
in the current unit of measurement in the form of
x.xxxxE±xx)