All steps from Normal operation to the different adjustment modes and related adjusting steps can also be found in
the following guide:
Adjustment User switch and LED-ring/LED action
Zero/FS
adjustment
Press button for 1 second until light ring is off
Release button to enter Zero/FS adjustment mode
Light ring is pulsing red and green
Press button >1 second to perform Zero/FS adjustment
Press button <1 second to
go back to Normal
Operation
LED-ring/LED is off
Release button
LED-ring/LED flashes green 10
times:
Zero/FS adjustment succeeded
LED-ring/LED flashes red 10
times:
Zero/FS adjustment failed
Setpoint 1
adjustment
Press button for 1 second until light ring is off
Continue pressing button for 3 seconds until light ring shows single green flash
Release button to enter Setpoint 1 adjustment
LED-ring/LED is pulsing green
Press button for >1 second
Press button <1
second to go back
to Normal
Operation
LED-ring/LED is off
Output current setpoint value
Continue pressing button for >5 seconds
Release button to
keep current voltage
output threshold to
Analog output and
return to Normal
operation
Ramp up/down voltage/setpoint value
Release button to keep current voltage output
threshold to Analog output
Tap button to toggle
through Output voltage
with small increments of
10 mV
Wait for >5 seconds
LED-ring/LED flashes
green
Wait for >5 seconds
LED-ring/LED flashes green 3 times
Selected Output voltage is stored
Setpoint 2
adjustment
Press button for 1 second until light ring is off
Continue pressing button for 6 seconds until light ring shows single red flash
Release button to enter Setpoint 2 adjustment
LED-ring/LED is pulsing red
Press button for >1 second
Press button <1
second to go back to
Normal Operation
LED-ring/LED is off
Output current setpoint value
Continue pressing button for >5 seconds
Release button to
keep current voltage
output threshold to
Analog output and
return to Normal
operation
Ramp up/down voltage/setpoint value
Release button to keep current voltage output
threshold to Analog output
Tap button to toggle
through Output voltage
with small increments
of 10 mV
Wait for >5 seconds
LED-ring/LED flashes
green
Wait for >5 seconds
LED-ring/LED flashes green 3 times
Selected Output voltage is stored
Содержание 230035V02