Object Structure
17200138_002_C0 – (2016-10) –
Leybold
19
Attr.
ID
Access
rule
NV/
V
Name DeviceNet
data type
Description of
Attribute
Semantics of Values
25 Set NV
Safe
State
USINT
Specifies the be-
havior for the
Value
for states
other than
Execute
→
"Semantics" section
below.
[default] = 0
26 Set NV
Safe
Value
INT or
specified
by
Data
Type
The Value to be
used for Safe
State = Safe
Value
→
"Semantics" section
below.
[default] = 0
32 Get NV
Over-
range
INT or
specified
by
Data
Type
Specifies the
highest valid
Value
The value above which
attribute
Reading Valid
is set to invalid.
[default] = maximum
allowable value for the
Data Type
33 Get NV
Under-
range
INT or
specified
by
Data
Type
Specifies the
lowest valid
Value
The value below which
attribute
Reading Valid
is set to invalid.
[default] = minimum
allowable value for the
Data Type
94 Get V
Sensor
Warning
Struct of
Byte
Bit definitions of
Sensor Warnings
0 = [default]
→
"Semantics" section
95 Get V
Sensor
Alarm
Struct of
Byte
Bit definitions of
Sensor Alarms
0 = [default]
→
"Semantics" section
96 Get V
Status
Extension
BYTE Bit-mapped
byte
providing addi-
tional status bits
Bit description:
0
Reading Invalid *)
1 Overrange
Exceeded
2 Underrange
Exceeded
99
Get
NV Subclass
UINT
Defines a subset
of additional attri-
butes, services
and behaviors.
02 = Heat Transfer
Vacuum Gauge
*) Logical inversion of
Reading Valid
.
2.4.2.2 Semantics of S-Analog Sensor Instance 1
All data type attributes use the enumerated values
integer or float
.
→
Abbreviations
4
The data type value will be set automatically based upon the first valid I/O con-
nection established by the device.
If no established I/O connections exist, which include an attribute from this object,
then the
Data Type
attribute is settable provided that the object is in the
Idle State
.
Note: Using data type integer in combination with a pressure unit (mbar, Torr or
Pa) will obviously not produce reasonable values below 1.
The Data Unit is only settable in the IDLE state.
ID 3
Data Type
ID 4
Data Unit