background image

10      

   

               

300360490_002_C0 - 09/2016 - © Leybold 

4.2 

Adjustment 

 
The transducer is adjusted ex works in upright position with 24V voltage supply. 
Through contamination, ageing or extreme climatic conditions the need for 
readjustment may arise. 
Readjustment on atmosphere or zero pressure is done digitally by means of the 
"up" and "down" buttons. The transducer will notice automatically which 
adjustment point is relevant.   

 
To achieve optimum results of the adjustment consider a 
warm-up of at least 10 minutes at the appropriate calibration 
pressure before any adjustment. 
 
 

4.2.1  Zero Adjustment 

The adjustment at zero pressure is an offset compensation of the measurement 
signal.  
 

For zero adjustment real pressure must be less than 0.1 mbar. 
An adjustment at zero pressure is only possible, if the DI2000 
pressure reading is below 100 mbar. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

The output signal at zero pressure should equal 4.000 mA (DI2000 / DI2001) or 
2,000 V (DU2000 / DU2001).  

 

For adjustment first remove the rubber caps from both buttons (1), then -by 
means of a screw driver or other suitable tool- push button "up" to increase the 
output signal or button "down" to decrease the output signal (2). If a button is 
permanently pressed the signal will change continuously in the respective 
direction. 
If no further button is pressed for an interval of 5s, the adjusted value is set. 
Now insert the rubber cap again. 

Содержание 15813V01

Страница 1: ...Vacuum Transducer DI2000 DI2001 DU2000 DU2001 Operating Instructions 300360490_002_C0 Cat No 15813V01 DI2000 Cat No 15815V01 DI2001 Cat No 230502V01 DU2000 Cat No 230503V01 DU2001...

Страница 2: ...ctrical Connection 7 3 3 1 Operation with Display Unit 7 3 3 2 Operation with Other Supply And Evaluation Units 8 4 Operation 9 4 1 General 9 4 2 Adjustment 10 4 2 1 Zero Adjustment 10 4 2 2 Zero Adju...

Страница 3: ...ed Consider possible reactions between materials and process media Consider possible reactions of the process media due to the heat generated by the product Do not carry out any unauthorized conversio...

Страница 4: ...put signal Unless these different features are described explicitly only model DI2000 is exemplarily mentioned in the following 2 2 Delivery Content Included in the delivery consignment are Transducer...

Страница 5: ...ers use EPDM as sealing material The transducer can be mounted to suitable flange connectors or by means of the G1 4 female thread Proper Use The DI2000 serves exclusively to provide total pressure me...

Страница 6: ...cleanliness and damage prevention when using vacuum compo nents Remove the protective cover is required again during maintenance work Make vacuum connection via small flange DN16 ISO KF If you use a...

Страница 7: ...d of small flange connec tions 2 5 bar abs Use clamps that are suitable for overpres sure operation 3 3 Electrical Connection 3 3 1 Operation with Display Unit The transducer can be connected to suita...

Страница 8: ...wn below DI2000 DI2001 Pin description connector topview Diode plug 7 pin male Pin 1 Pin 2 brown white brown Pin 3 blue white blue Pin 4 Pin 5 green white green Pin 6 orange white orange Pin 7 Schirm...

Страница 9: ...easure for the applied pressure Output Signal The analog output signal of the transducer is available immediately after the voltage supply has been connected The output of 4 0 20 mA of a DI2000 DI2001...

Страница 10: ...tment The adjustment at zero pressure is an offset compensation of the measurement signal For zero adjustment real pressure must be less than 0 1 mbar An adjustment at zero pressure is only possible i...

Страница 11: ...d be 0 mbar For adjustment first remove the rubber caps from both buttons 1 If the pressure reading is 0 in case of a DISPLAY unit or ccc0 cc0 or c0 in case of a CENTER unit the output signal is too l...

Страница 12: ...or an interval of 5s the adjusted value is set Now insert the rubber cap again 4 2 4 Full Scale Adjustment DU2000 DU2001 With Display Unit The full scale adjustment is a span correction of the measure...

Страница 13: ...and enclose a decontamination declaration The unit is not prepared for customer repair Malfunction of the unit which is caused by contamination wear and tear or a broken sensor is not covered by warra...

Страница 14: ...3 ceramic FPM DI2000 DU2000 EPDM DI2001 DU2001 Operating temperature 0 60 o C Storage temperature 40 70 o C Response time 20 ms Voltage supply DI 12 30 VDC two wire DU 14 5 30 VDC four wire Electrical...

Страница 15: ...300360490_002_C0 09 2016 Leybold 15 Declaration of Conformity...

Страница 16: ...Headquarter Leybold GmbH Bonner Strasse 498 D 50968 Cologne T 49 0 221 347 0 F 49 0 221 347 1250 info leybold com w w w l e y b o l d c o m...

Отзывы: