GA10.218_1.01 - 11/2004
6.8 Access Control
•
>
With this menu you can deny or allow access to specific functions of the
PhoeniX.
6.8.1 Access to CAL function
•
>
>
It can be selected whether the access to the calibration menu is restricted
or not.
6.8.2 Change Menu-PIN
•
>
>
The access to the menu can be restricted by entering or changing the
personal identification number (PIN). No PIN will be checked if 0000 is
entered.
The default setting for the Pin is 0013.
Please refer to Chapter
for the description of the
entry.,
Fig. 28 The Access Control Menu
Softkey 3
Access to Zero function
Softkey 4:
Softkey 8:
Softkey 3: Off
The calibration function is only available at the main menu. If the Menu-
PIN (See Chapter
) is activated you need this PIN to start a calibra-
tion.
Softkey 5: Help function
Softkey 7: ON
The calibration function is available at the main menu and in Stand-by and
the measure mode.