signal will be subtracted. If you press START when the value is not under-
lined, the old background signal from the last Stand-by will be subtracted.
6.2.6 Lower display limit
>
>
This mode limits the range (detection limit) of the measured leak rate. This
is valid for vacuum mode only.
6.3 Mode
>
The mode menu
enables the submenu to select the different wor-
king modes.
Softkey 3, 7: Changing of the lower detection limit between 1 x 10
-9
and
1 x 10
-12
The limit for the PhoeniXL300 dry ranges between 1 x 10
-9
and 1 x 10
-11
mbarl/s.
Softkey 5:
?
Help
Fig. 19 The Mode Menu
Key No. Name
Description
1
Cancel
Return to the main menu without any chan-
ges of the present settings.
3
Sniff
The sniffer mode will be used.
4
Not used in this menu.
5
Not used in this menu.
7
Vacuum
The vacuum mode will be used.
8
OK
Save the settings and return to the previous
menu.