Installation
20
⏐
GA10.218_1.01 - 11/2004
2.2.2 Connections for the Data Acquisition
Systems
2.2.2.1 Option (Accessories)
The sniffer line SL300 or the partial flow system may be connected to the
option port (Please refer to Chapter
/9):
Contact pins 1 and 3 are fused together with a 0.8 A slow-blow fuse. The
amount of power which can be drawn is limited to 10 W. The contacts are
numbered from bottom to top.
1. Vent
2. Exhaust
3. Gasballast
4. On / Off
5. Serial
6. Control 2
7. Record
8. Control
9. Option
10. Remote
Fig. 2
Pin
Assignment
1
+24 V, constantly applied, power supply for the LEY-
BOLD partial flow valve or sniffer line.
2
GND
3,
+24V switched by the PhoeniX for an external venting
valve
4, 5, 6, 7, 8
These pins are used in connection with accessories.