User Manual for LCO-i100 Console
www.levitronix.com
PL-4066-00, Rev03, DCO# 21-230
11
4.2 Electrical Installation
4.2.1 Connection Overview
shows the various interfaces of the console. The corresponding designations are listed in
Figure 9: Interfaces of console LCO-i100
Note 1: All connectors come delivered with a protective cap for a min. IP66 rated sealing if unused.
Note 2: When connected all connectors have a minimum rating of IP66.
#
Interface
Description
Specification/Purpose/Note
1
DC Supply Connector
Single-phase DC connection
24 VDC max. 6 A with external AC/DC supply
Software configurable Autoresume function.
1
RS485 for connecting PuraLev-i600SU with Y-split cable ICY-4.1.
2
2
Pump Driver
Power and signal connector for
IPD-30, IPD-100 drivers and IPD-600 drivers.
IPD-600 drivers can only be connected with a ICH-4.1 split cable.
2
3
Remote
1
RS485 for future purposes.
Not active. For internal usage.
4
PLC
Remote control.
Configurable designations.
2xdigital outputs (outputs software configurable to 24VDC
1
),
2xdigital inputs, 1xanalog output (4-20 mA), 1xanalog input (4-20mA)
Note: All signals isolated. See detailed description in
Section 4.2.2
5
Flow Sensor
LEVIFLOW
®
Connection of flow sensors from the
LFS-SU
and
LFSC-D
series.
Circular 6 pin snap-in connector.
Purpose: Flow control or monitoring.
6a
6b
6c
Pressure Sensor 1
Pressure Sensor 2
Pressure Sensor 3
1
Connection for single-use pressure sensors 1,
2 and 3.
Needs adaptor cable.
Circular plastic snap-in connector (on console side).
Purpose: pressure monitoring and pressure control.
7
Temperature Sensor
Connection for 1 single-use temperature
sensors.
Needs adaptor cable.
Circular plastic snap-in connector (on console side).
Purpose: Temperature monitoring.
8
Analog Input with 24 VDC Supply
Multi-purpose sensor input.
Current input 4-20 mA, 24 VDC supply (max. 0.5 A)
9
RS485 24VDC Out
1
Sensor input for clamp-on flow sensor series
LFSC-iX.
RS485 bus with 24 VDC supply.
10
LAN Interface
TCP/IP Modbus
For connection to OPC devices or
Levitronix
®
Service Software
.
11
USB Interface
Connection for a USB stick.
USB access for data debugging and software update through a USB stick.
12
Power Button
Switch on/off
Press shortly for controlled (confirmation by the user necessary) power off.
Press 4 s for immediate power off.
13
Touchscreen
Display and user touch interface.
7” multi-touch screen.
14
15
16
Type label
Serial number label
Certification (NRTL) label
--
--
Table 9: Interface specifications of console LPC-i100.1
Note 1: Not available on units older than 2018/19.
Note 2: Not available on units older than 2020.
14