25
ENGLISH
NOTE: The Tuning function is only available in alternate and static mode.
9.3.3 Therapy
The Therapy button allows you to choose one of the therapy modes. There are four selectable
therapy modes available. Continually press Therapy button to select a desired mode:
MaxFirm
Alternate
Static
Alt
Seat Inflate
Seat Inflate
A. Max Firm
The surface will go into Max firm automatically when the power switch is turned on. This
insures the pump is able to reach its maximum operating pressure. Once the max pressure
level is reached, the pump will automatically switch into alternating mode. User can also use
this function during patient transfers or nursing procedures for better support. This mode
will last for 20 minutes and then it will return to previous setting automatically. To disable
this function, press Maxfirm button again.
B. Alternate
The mattress will alternate every 10 minutes under this therapy mode.
C. Static
Non-alternating mode, all of the air cells is equally inflated.
D. Seat inflate
The seat inflate features additional supports to the patient during upright position.
9.3.4 Alarm Mute
Press alarm mute button to temporary suspend the low-pressure alarm (LED light and
buzzer). Should the situation not resolved within 5 mins, the alarm shall resume to notify
the patient.
A. PFA (Power Failure Alarm)
During power failure situation, the Power Failure LED light will lit on with buzzer.
By pressing the mute button to disable both buzzer and LED.
B. Low Pressure Indicator
When low pressure LED lights up, the pressure inside of air mattress is below normal.
Please refer to troubleshooting.
C. Tech. Support
This feature will light during mechanical failure situation. User can notify the technician for repair.
9.3.5 Panel Lock
If the control panel remains untouched for 5 minutes, a green LED will light and the control
panel will lock all settings. This will prevent the settings from being accidently changed
during normal operation. To unlock, simply press the Panel lock button for three seconds.