Display and control panel AMS 304i
Leuze electronic
AMS 304
i
40
TNT
35/7-2
4
V
8.2.2
LED status displays
PWR LED
Off
Device OFF
- No supply voltage
Flashing green
Power LED flashes green
- No measurement value output
- Voltage connected
- Self test running
- Initialization running
- Parameter download running
- Boot process running
Green continuous light
Power LED green
- AMS 304
i
ok
- Measurement value output
- Self test successfully finished
- Device monitoring active
Red flashing
Power LED flashes red
- Device ok but warning message (ATT, TMP, LSR)
set in display
- Light beam interruption
- Plausibility error (PLB)
Red
continuous light
Power LED red
- No measurement value output; for details, see Display
Orange
continuous light
Power LED orange
- Parameter enable active
- No data on the host interface
BUS LED
Off
BUS LED off
- No supply voltage (Power)
- PROFIBUS deactivated?- SSI interface active!