Operating Instructions

4

01056G

WARNING:  ALWAYS SHIELD EYES WHEN

WORKING NEAR BATTERIES.  DO NOT PUT
WRENCHES OR OTHER METAL OBJECTS
ACROSS BATTERY TERMINAL OR BATTERY
TOP.  ARCING OR EXPLOSION OF THE
BATTERY CAN RESULT!

Charger will turn ON one to three (1-3) seconds
after completion of the last of the above three
actions indicated by the glowing YELLOW light
and the charger ammeter indicating initial
charge current.

If the YELLOW light does not illuminate and the
Power switch is in the "ON" position, the AC
outlet is not live or an open circuit exits in the
wheelchairs internal circuitry.  Test for a live AC
outlet by connecting a suitable household
appliance to the specific outlet.  If the outlet tests
live, then the fault is in the wheelchair wiring.
Take the wheelchair to your dealer to have the
fault corrected (see Troubleshooting).

4.

Monitor ammeter for correct charge rate.  Initial
charge rate should be between 7 to 9 amps.  As
the battery reaches approximately 70% of full
charge, the charge rate decreases to a reading
determined by the "BATTERY TYPE" switch
setting and the condition of the batteries.

Replaceable Electrolyte Deep Cycle Batteries

With healthy batteries, the charging current will
decrease to between 2 to 4 amps and remain
there until the charger turns off.  As replaceable
electrolyte deep cycle batteries lose capacity,
the charge rate may no longer decrease to this
reading.  The charger will still determine when
the batteries are as charged as they are capable
of being and will turn off.  When battery capacity
has decreased to where the charger is no longer
able to taper down to less than 4 amps, or the
batteries will no longer perform as desired they
should be replaced.

Gel Cell and Sealed Deep Cycle Batteries

The charging current will decrease to near zero
and remain there until the charger turns off.  Gel
cell deep cycle batteries will still taper to near
zero charge current even as they wear out and
lose capacity and range.  As sealed deep cycle
batteries lose capacity and range, the charge
current may no longer decrease to this low a
reading.  The charger will still determine when
the batteries are as charged as they are capable
of being and turn off.  When the batteries will no
longer perform as desired, they should be
replaced.

Charger is equipped with a temperature sensor
which automatically turns charger off if it
overheats.  Should charger turn off before the

batteries are fully charged, check to be sure all
ventilation openings are free from obstructions.
After charger cools down to a safe temperature,
it will automatically restart.  If charger repeatedly
overheats, refer to a qualified service agent.

5.

Charger turns off automatically when batteries
are fully charged.  Charge time varies with
battery size and depth of discharge.  Allow 8
hours for normal charging.  Larger batteries
(greater than 85 ampere-hours) or severely
discharged batteries may require up to 14 hours
to be properly charged and equalized.  If the
charger operates for 14 hours and is unable to
fully charge the batteries, an internal timer turns
the charger off, extinguishes the YELLOW
charge light, and turns the RED fault light on.

6.

After the charger has turned off, move the
POWER switch to "OFF", disconnect AC cord
first, then the DC output cord from the
Wheelchair charging receptacle.

CAUTION:  TO AVOID DAMAGE TO THE

CHARGER CORD AND BATTERY CONNECTOR,
DISCONNECT BY GRASPING THE PLUG
HANDLE OR BODY AND PULLING IT STRAIGHT
OUT OF THE BATTERY CONNECTOR.  DO NOT
PULL ON THE CHARGER CORD.  DO NOT
TWIST, ROCK OR PULL THE PLUG SIDEWAYS.

WARNING:  TO PREVENT ARCING OR

BURNING NEAR THE BATTERIES, DO NOT
DISCONNECT THE OUTPUT CORD FROM THE
BATTERIES WHILE THE CHARGER IS
OPERATING.  IF THE CHARGE CYCLE MUST BE
INTERRUPTED, FIRST SLIDE THE POWER
SWITCH TO OFF, THEN DISCONNECT AC CORD,
AND FINALLY THE OUTPUT CORD FROM THE
BATTERIES.  KEEP SPARKS, FLAME, AND
SMOKING MATERIALS AWAY FROM
BATTERIES.  NO SMOKING.

WARNING:  DO NOT LEAVE THE DC

OUTPUT CONNECTOR PLUGGED IN WHILE
UNATTENDED FOR MORE THAN TWO (2) DAYS.
SEVERE OVERCHARGING AND DAMAGE TO
THE BATTERIES MAY RESULT IF THE
CHARGER DOES NOT TURN OFF.

PROPER CARE OF DEEP-CYCLE MOTIVE
POWER BATTERIES

Batteries used on wheelchairs are subjected to
severe deep-cycle duty on a daily basis.  For this
reason, it is important that only deep-cycle batteries
be used.  Although these batteries are designed to
withstand such duty, the following precautions must
be observed to obtain good performance and
maximum cycle life.

Содержание 13820 24EL8

Страница 1: ...racteristics are adjustable to charge gel cell and sealed or conventional replaceable liquid electrolyte deep cycle lead acid batteries Different charge characteristics are selected by a switch on the front of the charger Patented electronic circuit monitors battery state of charge and automatically turns charger off when the batteries reach full charge Line voltage compensation produces consisten...

Страница 2: ...n cooled design requires an unobstructed flow of cooling air for proper operation Keep all charger ventilation openings at least two inches 2 5 08cm away from walls and other objects Do not allow clothing blankets or other material to cover charger DO NOT place under bed WARNING CHARGERS CAN IGNITE FLAMMABLE MATERIALS AND VAPORS DO NOT USE NEAR FUELS GRAIN DUST SOLVENTS THINNERS OR OTHER FLAMMABLE...

Страница 3: ... cord is disconnected CAUTION MAKE SURE THE BATTERY PACK IS A 24 VOLT 12 CELL SERIES CONNECTED 28 85 AMP HOUR 20 HR RATE RECHARGEABLE DEEP CYCLE BATTERY SYSTEM THE GEL CELL AND SEALED DEEP CYCLE BATTERY SETTING CHARGES TO A MAXIMUM ON CHARGE VOLTAGE OF 2 37 VOLTS PER CELL THE REPLACEABLE ELECTROLYTE DEEP CYCLE BATTERY SETTING HAS AN ON CHARGE VOLTAGE OF AT LEAST 2 50 VOLTS PER CELL VISUALLY CHECK ...

Страница 4: ...es are as charged as they are capable of being and turn off When the batteries will no longer perform as desired they should be replaced Charger is equipped with a temperature sensor which automatically turns charger off if it overheats Should charger turn off before the batteries are fully charged check to be sure all ventilation openings are free from obstructions After charger cools down to a s...

Страница 5: ...ent excessive self discharge Keep battery terminals reasonably tight PERSONAL PRECAUTIONS 1 Someone should be within range of your voice or close enough to come to your aid when you work near a lead acid battery 2 Have plenty of fresh water and soap nearby in case battery acid contacts skin clothing or eyes 3 Wear complete eye protection and clothing protection Avoid touching eyes while working ne...

Страница 6: ...d to take place it may be partially reduced by returning temporarily to daily charging The decrease in wheelchair range where the chair loses power earlier and earlier in the day indicates one of the following a The electrolyte level in conventional liquid electrolyte lead acid batteries was allowed to drop below the top of the battery plates Add distilled water to just cover the tops of the plate...

Страница 7: ...1 CASE ASSEMBLY 16246S 1 TRANSFORMER ASSEMBLY 12245S 1 ELECTRONIC TIMER ASSEMBLY 13811S 1 AMMETER 10 AMP 14308S 1 CONTROL CABLE ASSEMBLY 15184S 1 SWITCH SLIDE SPST 125 VAC 6 AMP 14341S 1 THYRISTOR ASSEMBLY 04275S 2 BUSHING INSULATOR 6N3 4 FOR CORDSETS 14444S 1 CORDSET AC 18 3 102 14421S 1 CORDSET DC 14 2 108 BLACK RED POWERPOLE 14422S 1 CORDSET DC 14 2 108 YELLOW WHITE POWERPOLE 14538S 1 CORDSET D...

Страница 8: ... completed with a reasonable time after receipt of defective items This warranty does not cover any semiconductor parts such as diodes which are vulnerable to electrical overloads beyond the control of Lester Warranty on parts not manufactured by Lester which include but are not limited to timers and ammeters is limited to the period specified in the original manufacturer s warranty This warranty ...

Отзывы: