Chapter 25 - Guidance and Declaration for Electromagnetic Compatibility
The
Lenire
system has been tested for electromagnetic performance to
the following standards:
»
IEC61000-4-3 (radiated electromagnetic fields)
• 80 MHz to 2.7 GHz at 10V/m
• 2.7 GHz to 6 GHz at 3V/m
• Spot frequencies per EN 60601-1-2:2015 Table 9 at up to 28 V/m
»
IEC 61000-4-2 (Electrostatic Discharge)
• +/- 8kV Contact Discharge
• +/- 15kV Air Discharge
»
IEC 61000-4-8 (50Hz/60Hz magnetic fields) at 30 A/m
Disturbances higher than these limits are unlikely to be encountered in the
home healthcare or professional healthcare environments.