Positioning the Cuff 

 
 

1.

 

Remove all accessories (watch, bracelet, etc.) from your left wrist. If your 

physician has diagnosed you with poor circulation in your left wrist, use your 
right wrist. 

2.

 

Roll or push up your sleeve to expose the skin. 

3.

 

Apply the cuff to your left wrist with your palm facing up. 

4.

 

Position the edge of the cuff about 1-2 cm below the wrist. 

5.

 

Fasten the wrist cuff around your wrist, leaving no extra room between the cuff 

and your skin. If the cuff is too loose, the measurement will not be accurate. 

6.

 

Patients with Hypertension: 

The middle of the cuff should be at the level of the right atrium of the heart; 
 
Before starting measurement, please sit comfortably with legs uncrossed, feet flat 
on the floor, back and arm supported. 
 

• Rest for 5 minutes before measuring. 
• Wait at least 3 minutes between measurements. This allows your blood 
circulation to recover. 
• For a useful comparison, try to measure under similar conditions. For 
example, take daily measurements at approximately the same time, on the same 
wrist, or as directed by a physician. 
 

 

 
 
 

Содержание Sharon LH67402

Страница 1: ...User s Manual LEICKE Sharon Blood Pressure Monitor with Bluetooth Product number LH67402 ...

Страница 2: ...our company s efforts We take your reviews and proposals seriously and evaluate them continuously That way we get to know you and your demands on our products and services better and thus allow for positive enhancements To discover more about LEICKE products or if you have any questions about this product visit our website www leicke com PACKAGE CONTENTS Before attempting to use this device please...

Страница 3: ... C to 40 C Relative humidity 85 RH Atmospheric pressure 86kPa to 106kPa Storage transportation condition Temperature 20 C to 60 C Relative humidity 10 RH to 93 RH Atmospheric pressure 50kPa to 106kPa Degree of protection Type BF applied part Protection against ingress of water IP22 Software version V01 Device classification Battery Powered Mode Internally Powered ME Equipment AC Adaptor charged Mo...

Страница 4: ...Memory Up to 60 pieces of records MEASUREMENT PRINCIPLE This product uses the Oscillometric Measuring Method to detect blood pressure Before every measurement the unit establishes a zero point equivalent to the atmospheric pressure Then it starts inflating the cuff Meanwhile the unit detects pressure oscillation generated by beat to beat pulsatile which is used to determine the systolic pressure a...

Страница 5: ...exist Check with your local authority or retailer for recycling advice Symbol for MANUFACTURER Symbol for SERIAL NUMBER Symbol for DIRECT CURRENT Symbol for Including RF transmitter Symbol for Authorised Representative in the European Community Caution These notes must be observed to prevent any damage to the device Symbol for MANUFACTURE DATE INDICATIONS FOR USE The LEICKE Sharon Blood Pressure M...

Страница 6: ...ygen nitrous oxide Please keep the unit out of reach of infants children or pets since inhalation or swallowing of small parts is dangerous or even fatal Please use ACCESSORIES and detachable parts specified authorised by MANUFACTURER Otherwise it may cause damage to the unit or endanger the user patient The patient is an intended operator The patient can measure transmit data and charge battery u...

Страница 7: ...ressure Pulse Beat minute Low Battery Low battery please charge mmHg Unit Measurement unit of blood pressure 1mmHg 0 133kPa IHB Detector Irregular Heartbeat Detector Bluetooth Successful Bluetooth Connection Error Error Memory Recalling the history records Time Hour Minute Month Day Year Heartbeat Heartbeat detection during the measurement ...

Страница 8: ...onents SYSTOLIC DIASTOLIC PULSE RATE TIME MEM UP START STOP BUTTON BUTTON SET DOWN BUTTON Component list of pressure measuring system CUFF Type BF applied part 1 PCBA 2 Air Pipe 3 Pump 4 Valve 5 Cuff USB Interface ...

Страница 9: ...semble yourself or by unauthorized maintenance personnel 2 When used normally it can be charged about 300 times if the battery cannot be charged normally or the blood pressure monitor cannot be used normally please contact authorized maintenance personnel If you measure three times per day and the battery was fully charged it can be used for about 20 days 3 Store and use the blood pressure monitor...

Страница 10: ...empt to disassemble the blood pressure monitor or force open the built in battery 12 Do not clean the blood pressure monitor when it is being charged Always unplug the charger first before cleaning the blood pressure monitor 13 Do not dispose of your blood pressure monitor in a fire The battery could explode causing injury or death 14 Batteries battery pack or batteries installed shall not be expo...

Страница 11: ... Time Format 12 Hours 1 When the monitor is OFF press and hold SET button for 3 seconds to enter Time Setting Mode 2 As pictured in the right the blinking number represents the HOUR Press MEM button to change the number Each press will increase the number by one in a cycling manner 3 Press SET button again to confirm HOUR Then the number representing MINUTE blinks ...

Страница 12: ...Setting Date and Time 4 Repeat steps 2 and 3 to confirm MINUTE 5 Repeat steps 2 and 3 to confirm MONTH DAY and YEAR 6 After confirming YEAR the LCD will display dONE and then shut off ...

Страница 13: ...etween the cuff and your skin If the cuff is too loose the measurement will not be accurate 6 Patients with Hypertension The middle of the cuff should be at the level of the right atrium of the heart Before starting measurement please sit comfortably with legs uncrossed feet flat on the floor back and arm supported Rest for 5 minutes before measuring Wait at least 3 minutes between measurements Th...

Страница 14: ...uetooth and the app Make sure both are ON when pairing is proceeding 2 When the monitor is OFF press and hold the START STOP button to start pairing These symbols will be shown on the LCD alternatively indicating pairing is proceeding If SUCCEED this symbol will be shown on the LCD If FAIL than this symbol Bluetooth ERROR will be shown on the LCD 3 The monitor will shut off after pairing process i...

Страница 15: ...ement 1 After correctly positioning the cuff press START STOP button to turn on the monitor and it will complete the measurement process Adjust to zero Inflating and measuring Display and save the measuring result ...

Страница 16: ...on the LCD indicates data transmission 3 If the data is successfully transmitted the LCD will display dONE If data transmission fails the LCD will display BLUETOOTH ERROR instead 4 Press START STOP button to turn off the monitor Otherwise if there is no operation it will switch off within 1 minute ...

Страница 17: ...dings first 2 Press MEM UP button or SET DOWN button to rotate the history records MEM UP to go forward SET DOWN to go backward CAUTION The most recent record 1 is shown first Each new measurement is assigned to the first 1 record All other records are pushed back one digit e g 2 becomes 3 and so on and the last record 60 is dropped from the list ...

Страница 18: ...e MEM button and the SET button for 3 seconds 2 The LCD will display dEL dONE indicating that memory clearing is complete And then it will shut down CAUTION Under Memory Recalling Mode if you wish to give up clearing press START STOP to turn off the monitor 3 When there is no memory in the monitor if you press the MEM button to look up history the LCD will display as pictured to the right ...

Страница 19: ...f taken in the following circumstances Within 1 hour after dinner or drinking Immediate measurement after tea coffee smoking Within 20 minutes after taking a bath When talking or moving your fingers In a very cold environment When you need to pee urgently ...

Страница 20: ...etting up maintaining or using please contact the customer service Don t open or repair the device by yourself 3 Please report to LEICKE if any unexpected operation or events occur 4 Cleaning Dust environment may affect the performance of the unit Please use a soft cloth to remove dirt from the device and cuff before and after use 5 Calibration The manufacturer does not require such preventive ins...

Страница 21: ...value in the cycle which is called diastolic pressure What is the standard blood pressure classification The blood pressure classification published by World Health Organization WHO and International Society of Hypertension ISH in 1999 is as follows CAUTION Only a physician can tell your normal BP range Please contact a physician if your measuring result falls out of the range Please note that onl...

Страница 22: ... a pulse irregularity consistent with an irregular heartbeat was detected during measurement Usually this is NOT a cause for concern However if the symbol appears often we recommend you seek medical advice Please note that the device does not replace a cardiac examination but serves to detect pulse irregularities at an early stage Why does my blood pressure fluctuate throughout the day Individual ...

Страница 23: ...d to pay attention to when you measure your blood pressure at home If the cuff is tied properly If the cuff is too tight or too loose If the cuff is tied on the wrist If you feel anxious Taking 2 3 deep breaths before beginning will be better for measuring Advice Relax yourself for 4 5 minutes until you calm down Is the result the same if measuring on the right wrist It is ok for both wrists but t...

Страница 24: ...hed on or within the distance range 1 shows Inflation is slow or the cuff is not secure Refasten the cuff and then measure again 2 shows The cuff is very tight Refasten the cuff and then measure again 3 shows Refasten the cuff and then measure again The pressure of the cuff is too high 10 or 11 shows Relax for a moment and then measure again The monitor detected motion talking or the pulse is too ...

Страница 25: ...s for basic safety and essential performance Collateral standard Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment Electromagnetic compatibility IEC EN 60601 1 2 2007 Medical electrical equipment Part 1 2 General requirements for basic safety and essential performance Collateral standard Electromagnetic compatibility Requirements a...

Страница 26: ...below The customer or the user of the device should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 2 The device must emit electromagnetic energy in order to perform its intended function Nearby electronic equipment may be affected RF emissions CISPR 11 Class B Harmonic emissions IEC 61000 3 2 Not applicable V...

Страница 27: ...spital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT...

Страница 28: ...recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all si...

Страница 29: ...the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 167 P 80 MHz to 800 MHz d 1 167 P 800 MHz to 2 5 GHz d 2 333 P 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 690 3 690 7 378 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the reco...

Страница 30: ...a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help This device complies with Part 15 of the FCC Rules Operation is subject to the following conditions This device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation FCC Caut...

Страница 31: ...CONTACT If you have any questions please feel free to contact us PRODUCTION WHOLESALE LEICKE GmbH Dohnanyistr 28 04103 Leipzig Germany Web www leicke com Mail info leicke com ...

Отзывы: