82
TS11/TS15, Check & Adjust
If
Calibrate the automatic target aiming
is checked and an ATR is available,
the adjustment will include the determination of the ATR Hz and V adjust-
ment errors.
)
Use a clean Leica standard prism as the target. Do not use a 360°
prism.
5.
Aim the telescope accurately at
a target at about 100 m distant.
The target must be positioned
within ±9°/±10 gon of the hori-
zontal plane.
The procedure can be started in
any face.
Step
Description
± 9°
~ 100 m
TS_069