Depending on the work situation, it
may be necessary to restrict the maxi-
mum travel path of the stereomicroscope. This
prevents the following:
Å
Injuries when manipulating the specimen
because the fingers or hand get pinched.
Å
Accidental contact between the objective
and the specimen, and potential resulting
damage.
Readjusting the motorized focus
The motorized focus is factory-adjusted and
normally does not need to be readjusted – even
if the maximum travel path is changed.
•
Exception: If the power fails while the moto-
rized focus is moving, the position data
are lost. In this case, the calibration must be
repeated using the Leica LAS software or the
Leica SmartTouch™. To do so, please consult the
respective manual.
Restricting the bottom travel range
1. Move the motorized focus into the lowest
position you want to reach.
2. Unscrew the screw of the limit stop on the
side of the focusing column.
3. Push the limit stop to the height of the
motorized focus.
It is easiest to move the limit stop by keeping the
screwdriver inserted and moving it upwards.
4. Tighten the screw of the limit stop.
Motorized Focus: Restricting the Travel Path
Содержание TL5000 Ergo
Страница 1: ...Living up to Life User Manual Leica TL5000 Ergo Transmitted Light Base ...
Страница 7: ...Leica TL5000 Ergo User Manual 7 Leica TL5000 Ergo ...
Страница 11: ...Leica TL5000 Ergo User Manual 11 Assembly ...
Страница 19: ...Leica TL5000 Ergo User Manual 19 Quick Start Guide ...
Страница 33: ...Leica TL5000 Ergo User Manual 33 Preparation for the Automatic Control System ...
Страница 38: ...Leica TL5000 Ergo User Manual 38 Operation with Leica SmartTouch ...
Страница 49: ...Leica TL5000 Ergo User Manual 49 Dimensions and Specifications ...
Страница 52: ...Leica TL5000 Ergo User Manual 52 Appendix ...