53
SMD Components
9.1.3
Application area FCS/FCCS, FLIM, FLCS, gated FCS
9.1.3.1
Special SMD Hardware
Table 6:
Special SMD hardware for FCS/FCCS, FLIM, FLCS, gated FCS
9.1.3.2
Software License
Table 7:
Software license for FCS/FCCS, FLIM, FLCS, gated FCS
9.2
Beam Path
The fluorescence necessary for SMD measurement is excited by means of lasers, which
are coupled into the scan head. For FCS and FCCS, continuous wave (cw) lasers are
normally used. However, pulsed lasers can also be used. For FLIM, FLCS or gated FCS, only
pulsed lasers can be used.
The excitation light reaches the specimen through the scan head. The light is collected by
the objective and passes through the scan head to the SMD detectors that can be either
within the spectral module in the scan head or outside the scan head.
For SMD, special detectors are needed that work in single photon counting mode and have
a low background signal. For FLIM, additionally, a high time resolution is necessary,
whereas for FCS a high quantum efficiency is crucial. Therefore, different detectors can be
used, depending on the field of application (see
and
).
When confocal, external detectors are used, light is split off to optical fibers at the external
interface X1 ("X1 port") with the aid of the X1 adapter, then guided to the photodetectors via
multimode waveguides (
). These waveguide fibers enable the scan head to be
mechanically decoupled from the detectors. If two external detectors are used, light is split
by a beam splitter into the two detection channels. The beam splitter is situated in the SMD
filter cube that is placed in the X1 port adapter on the X1 port.
Pulsed laser laser
coupling unit (LCU) + coupling
as with MP-FLIM
Detectors for SMD
External detection:
PDM APD from MPD ("MPD") + DSN102
And/or internal detection:
SP FLIM photomultiplier (within spectral module, used
for spectral FLIM)
Counting unit
TCSPC unit PH300 + Router PHR800
Leica LAS AF
FCS FLIM Wizard
PQ SPT
SPT1+2
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......