211
Summarized manual for FCS or other point measurements
6. Select the measurement times.
7. Start the FCS measurement in LAS AF by selecting the image acquisition button
Run FCS
under the step
Measurements
.
Normally, the SymPhoTime software will start the data acquisition automatically. If you
start the FCS data acquisition by selecting
Acquire Point Measurements
in the
SymPhoTime software, the measurement is not synchronized with LAS AF. The file
name, comments and instrument parameter settings are not transferred from LAS AF to
SymPhoTime.
8. The Measurement Preview dialog opens in SymPhoTime on the
FCS
tab (
left
).
The FCS curve is integrated over time.
Figure 183:
Working area of the annotation file
9. At the end of the measurement, the result can be viewed in the annotation file which can
be opened from the workspace (
). Open the
Correlation
tab as shown in
. The curves can be analyzed by selecting the
FIT
button that corresponds to
the cross-correlation or autocorrelation curve.
10. Starting from the acquired raw data file (in this case,
FCS 01.pt3
), the FCS curve can be
recalculated with enhanced functions, such as Time Gated FCS or Fluorescence Lifetime
Correlation Spectroscopy (FLCS) using the
FCS trace
button. The prerequisite for both of
these modes is that the FCS measurement was carried out using pulsed lasers for
excitation.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......