F(L)CS Data Acquisition
18.3.5
Optimizing FCS Settings
1. First, use the image display to find a measuring point in the image. Move the crosshair in
the image display to the desired position (
Figure 175:
Selecting the measuring points
2. Press the
Run FCS Test
button (
) on the
Setup
tab to start the
illumination. You can view and optimize the fluorescence intensity in the count rate
monitor. Laser line and intensity can be modified during the test measurement until you
attain the count rate or counts per molecule desired.
3. The scanner cannot be moved out of its defined position while
Run FCS Test
is still active.
To move the scanner to another position, you first have to stop the FCS measurement by
clicking the
Stop FCS Test
button.
4. Then, move the crosshair to the desired position and restart the measurement by clicking
the
Run FCS Test
button.
When you start the test measurement, an online correlation curve is displayed in
SymPhoTime.
A scanning operation is required before an FCS measurement
After starting LAS AF, at least one scanning operation has to be carried
out (using
Live
or
Capture
) before an FCS measurement can be started.
This is also necessary when working in solutions only. Otherwise, no FCS
measurement is started.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......