144
FLIM Data Acquisition
2. Click the upper radio button (
).
3. Filter selection:
•
For automatic filter selection: Select the
Auto Select
) check box.
Individual filters now cannot be manually selected.
•
For manual filter selection: Remove the check in the
Auto Select
check box. From the
filter list, select the suitable filter.
16.2.3.2
Setting for External FLIM or Intensity Image Acquisition
For external FLIM detection or normal intensity image acquisition, it is possible to use the
Empty
position (
)
as needed in the
Fluorifier Disc Settings
. Remove the
check in the
Auto Select
check box and select
Empty
in both menus.
Alternatively, if the effect of polarization is to be investigated, the lower radio button can be
clicked (
) and the polarization direction defined by moving the slider
position.
16.2.3.3
Setting for FLIM White Light Laser
If the white light laser is used as an excitation source, it is recommended - independent of
the detectors used - to use the filters of the fluorifier disc and, thereby, to select the
corresponding laser wavelengths (usually 488, 514, 561, 594, and 633 nm,
Figure 127:
Selecting the laser lines for FLIM white light lasers
Selection switches to set the laser wavelength and the filter wheel can be found in the
Setup FLIM
and
Measurements
steps on the
Acquisition tab (Figure 128
o
Figure 128:
Selection switches to set the laser wavelength and the filter wheel when using the white light
laser
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......