130
FLIM Data Acquisition
3. Deactivate all PMTs.
4. Activate the APDs by clicking the corresponding buttons (
).
5. Select low laser intensities.
6. You may also have to adapt the color look-up table (LUT). In the dialog of the respective
detector, click the color scale to call up the
LUT Selection
dialog. There, you can select
color look-up tables (LUT) for displaying the images to be acquired and create user-
defined color look-up tables.
You can also configure the intensity setting via
Gain
and the control panel. This function
has no effect on the detector; it modifies the color look-up table, which renders low-
intensity structures more visible. It may also be helpful to accumulate over several
images. The
Offset
setting on the control panel has no effect on the image.
7. Optimize the image quality. The functions for this are the same as outside the wizard.
8. Acquire an image by clicking the
Capture Image
image acquisition button or acquire a
stack by clicking the
Start
image acquisition button. The start and finish of the stack are
defined interactively. The stack image is automatically added to your LAS AF experiment.
16.1.1.3
External FLIM Photomultiplier
1. For light to fall on the detector, select the
---
option for
X1 Port
.
2. You can enable the
PMT Extern
option in
Beam Path Settings
. It is required for
automated functions in the FLIM Wizard (
Setup FLIM
and
Measurements
steps).
Activating it causes empty images to be created in LAS AF.
16.1.2
Selecting Laser Lines as an Excitation Source for Image Acquisition
All available lasers can be used for the image acquisition. You will find a description of the
mode of operation in
16.1.2.1
Using Continuous Wave Lasers
1. For capturing the emitted fluorescence, set the acousto-optical beam splitter to
fluorescence (
).
Dependency of the spectral detection range of the APDs
The spectral detection range of the APDs is determined by the SMD filter
cube used, not by the slider settings for the photomultipliers.
Images of the external FLIM PMT only in SymPhoTime
The signal of the external FLIM PMT is not displayed in LAS AF. You can
view images created by this detector only in SymPhoTime. The
corresponding detector displayed in LAS AF is a simulator required for
the function of the wizard.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......