131
FLIM Data Acquisition
Figure 116:
Dialog for AOBS configuration
2. Choose a suitable laser line and adjust the intensity to the desired level using the AOTF
slider.
3. If the system has a multifunction port (
MFP
), set it to
Substrate
.
16.1.2.2
Using MP Lasers
1. For MP scanning operations, select the desired wavelength in
Beam Path Settings
.
2. Choose a suitable neutral density filter for the attenuation using the
Trans
slider and/or
adjust the electro-optical modulator (EOM) using the
Gain
and
Offset
sliders.
3. If the system has a multifunction port, it should be set to
SP680
or
SP700
.
For more detailed information about switching on the MP lasers, refer to
.
16.1.2.3
Using Pulsed VIS Lasers
1. Close the shutter of the continuous wave VIS laser in
Beam Path Settings
and move all
sliders to the 0 position.
2. Open the shutter for pulsed lasers by activating the corresponding button (
).
3. Activate the desired laser by moving the corresponding slider upwards (
).
Prevent light incidence during MP FLIM measurement
During an MP FLIM measurement (particularly with HyD RLD), the light in
the room should be switched off and the shutter of the fluorescence lamp
should be closed. Furthermore, the iris diaphragm on the condenser
should be completely closed.
Controlling laser intensity of the pulsed laser
For pulsed lasers, the slider does not affect the laser intensity. Laser
intensity needs to be controlled at the desired laser.
Содержание TCS SP8 SMD
Страница 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Страница 4: ...4 Copyright ...
Страница 14: ...14 Contents ...
Страница 18: ...18 Intended Use ...
Страница 20: ...20 Liability and Warranty ...
Страница 28: ...28 General Safety Notes ...
Страница 32: ...32 Additional Notes on Handling the System ...
Страница 44: ...44 System Overview and Properties ...
Страница 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Страница 80: ...80 Safety Features ...
Страница 102: ...102 Switching On the System ...
Страница 116: ...116 LAS AF ...
Страница 214: ...214 Changing the Specimen ...
Страница 216: ...216 Changing the Objective ...
Страница 218: ...218 Piezo Focus on an Upright Microscope Figure 186 Piezo focus controller Figure 187 Spacer on objective ...
Страница 238: ...238 Switching Off the System ...
Страница 242: ...242 Repairs and Service Work ...
Страница 244: ...244 Maintenance ...
Страница 246: ...246 Disassembly and Transport ...
Страница 248: ...248 Disposal ...
Страница 254: ...254 Contact ...
Страница 256: ...256 Recommended Literature ...
Страница 266: ...266 Appendix Figure 225 Declaration of conformity ...
Страница 268: ...268 Appendix ...
Страница 269: ......