Leica L5 FL

User manual

Содержание L5 FL

Страница 1: ...Leica L5 FL User manual...

Страница 2: ...omicroscopes M series 13 Stereomicroscopes S series 14 Remote control connection 15 Lamp Installation replacement 16 Starting up 17 Luminous intensity setting 18 Electronic setting 18 Mechanical diaph...

Страница 3: ...handle 10 Lamp housing with halogen cold light reflector lamp 11 Release button for lamp housing 12 Power supply cord 13 Ventilation openings front side of instrument 14 Ventilation openings side of...

Страница 4: ...ooms only see page 25 Technical Data This instrument is not intended for operation in explosion prone areas A safe disconnect from the power supply system is achieved by disconnecting the power plug T...

Страница 5: ...teristic of converting incident light into heat damages may occur to heat sensitive or inflammable light absorbing materials To avoid such thermal damages and possible fire hazards or risk of burns pl...

Страница 6: ...e illumination to the absolute necessary measure It is absolutely necessary that the Leica L5 FL cold light source is operated with the voltage listed on the identification label 17 all ventilation op...

Страница 7: ...el at risk Caution observe the documentation Nonobservance can cause malfunctioning or damage to the equipment Warning about hot surfaces Risk of burns in case of a hot surface Important Information T...

Страница 8: ...8 Leica Stereomicroscope MZ6 with Leica L5 FL Leica Stereomicroscope S8 APO with Leica L5 FL...

Страница 9: ...itions and remove it The adapter is inserted in reverse order Please store the connecting adapter in a safe place for later use with other fiber optic light guides Screw the black ring back on the fib...

Страница 10: ...atures position marks the position 0 through 4 are marked by means of raised points in corresponding numbers The excitation filter is inserted in the Leica L5 FL cold light source To insert the filter...

Страница 11: ...The daylight filter is part of the standard delivery and is used to increase the color temperature for bright field applications The filter wheel features 5 positions for max five filters Fit the fil...

Страница 12: ...freely movable The support arm can now be positioned Next turn the rotary knob clockwise to lock all joints Hinged arm for illumination and UV protective filter Stereomicroscopes M series Connection o...

Страница 13: ...suppression filters The standard equipment contains one suppression filter The second position can be used for an additional suppression filter Filter slide housing with filter slide and filters One...

Страница 14: ...required so that the working distance remains unchanged if the user switches between fluorescence and bright field application The filter slide housing is mounted on the stereomicroscope as shown in F...

Страница 15: ...ntensity control 3 is deactivated Remote control and tilt switch allow switching the light source to STANDBY mode to protect the halogen lamp A corresponding note 21 appears on the display Please alwa...

Страница 16: ...housing 10 by pressing the release button 11 and pull it out to the stop Fig 1 Press both levers of the special socket down and pull out the defective bulb The two levers must also be pressed down whe...

Страница 17: ...ec ted fiber optic light guides Position The instrument is switched off Position The instrument is switched on The Leica L5 FL is equipped with a soft start to protect the halogen lamp that reduces an...

Страница 18: ...so results in a change of the color temperature of the emitted light The mechanical adjustment allows for the continuous change of the luminous intensity with constant color temperature Four lock in p...

Страница 19: ...lb manufacturer type dependent The LCD display 5 shows the approximate color temperature of the light emitted by the halogen lamp 18 The color temperature of the light can be adjusted by turning the l...

Страница 20: ...y opened diaphragm achieves the maximum brightness Turning the adjusting knob from one lock in position to the next results in approximately 2x increase or 1 2x decrease of the luminous intensity The...

Страница 21: ...en dent upon the magnification and the working distance of the stereomicroscope The focusing lens features a slide mechanism to obtain a perfect adjustment of the light spot diameter to the field of v...

Страница 22: ...Use a soft dry cloth or commercial plastic clean sing wipes to clean the outside of the instrument The display features an error status display 19 Malfunctions can be detected quickly and easily Err...

Страница 23: ...ment with switched off light source Lamp off fan operating Bulb defective Replace bulb see p 16 error status display No bulb Insert with light source Err1 switched off Lamp off fan operating Insuffici...

Страница 24: ...bulb type that will allow for perfect light utilization and illuminated area is deliver ed The following functions can be controlled via remote control Light on off STANDBY Electronic luminous intens...

Страница 25: ...e C 5 40 Relative humidity up to 31 C ambient temperature 85 from 31 C to 40 C ambient temperature linear decrease to 75 Atmospheric pressure hPa 700 1060 Transport and storage Temperature C 40 70 Rel...