![Leica DMI Series Скачать руководство пользователя страница 97](http://html1.mh-extra.com/html/leica/dmi-series/dmi-series_instructions-manual_1873304097.webp)
97
8. Operation
8.7 Port selection
Leica DMI 3000 B and Leica DMI4000 B:
Manual shifter rod activates and deactivates the
left-hand photo port.
VIS LEFT
100%
0%
20% 80%
alternatively: 0%
100%
Leica DMI 6000 B:
The
button on the front control panel switches 100%
of the light to the eyepieces.
Use the
button, also on the front control panel, to select
the side ports.
8.6 Tubes
Note:
Close any unused tube openings, as otherwise
stray light can interfere with observation.
Adjusting the viewing distance
• Adjust the viewing distance of the eyepiec-
es so that a congruent total image is seen
(Fig. 115).
Adjusting the viewing angle
• Ergotubes feature a tilting binocular section for
a 30–45° viewing angle adjustment range.
Beam splitting in photo tubes
The beam splitting is set manually by pulling out
a control bar.
Observation
photo
100% 0%
0% 100%
alternatively 50%
50%
BL
activation of Bertrand lens*
Light distribution via
manual control bar
Fig. 115
Tube
setting
♦→
→
→