79
HistoCore SPECTRA CV
Daily Instrument Setup
6
1
2
Fig. 55
Note
New processing in the
L2
P. 122 – 7.3.2 Extended Prime)
coverslip line is only possible after the
insertion of a new mounting medium bottle and the necessary extended priming.
To change the mounting medium bottle, proceed as follows:
1. Open the hood
.
2. Grasp the bottle sledge on the handle and move it forwards
3. Carefully withdraw the cannula
L2
from the mounting medium bottle opening and
place in the parking location
.
4. Remove the empty mounting medium bottle
and dispose of it in accordance with
the laboratory specifications.
1
2
4
3
Fig. 56
5. Take a new mounting medium bottle
out of the packaging and remove the black
Warning
Do not remove the white protective foil
).
It must remain on the bottle.