Version 1.3, Revision D
Operation
5
Program list panel
Fig. 64
1
2
3
Program list panel is displayed when no program is running
. It is switched to the program
running panel after a selected program starts
.
If a program is selected, the program icon and name is highlighted
Maximum 5 favorite programs can be shown in the program list panel. For how to set a program as a
favorite program, see
P. 72 – 5.4.2 Favorite programs)
5.6
Reagent handling
Caution
Always replenish reagents via the reagent bottle and paraffin via the paraffin bath.
Do not add reagents or paraffin directly into the retort. Doing so will cause an overflow from reagent
bottles or paraffin baths during filling or draining which could result in instrument malfunction or potentially
a fire accident.
5.6.1 Filling/draining reagents
Caution
The following steps should only be carried out by trained laboratory personnel, experienced in handling
reagents.
Always wear chemical protective clothing, protective goggles, respirator, rubber gloves and all other
necessary personal protective equipment, when handling the reagents used in this instrument.
Reagents used for tissue processing can be both toxic and/or flammable.
Caution
Dispose of waste reagent with care according to local regulations and the waste management policy of
the company or institution.