background image

Doc No. FM0851 issue A Page 6  

Registering a Heat Detector onto a

 

Touchsafe Pro 

Nursecall system

 

 

1. Put the Touchsafe Pro Master Panel into Telecare registration mode; touch the screen and Logon as an Engineer or a Manager 

(password required), 

then from the main menu page touch the 

Telecare

 

button to open the screen shown below…

 

 
 

 

 

 

 

 

 

 
 
 
 

 

 

 

 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

2. Enter the heat detector 

ID…

 

   

 

 
 
 

 

 

 

 

 

 

 

 

 
 
3. Press the 

Close

 button to exit Telecare mode and the Nursecall panel is now back in normal operating mode. 

 
4. Activate the heat detector and check the call is 

displayed on the Nursecall panel like this…

  

 
 
 
 
 
 
The screen banner is 

red

 for a heat alarm call (

Touch the banner to display notes

). 

A heat alarm call cannot be accepted or silenced from the screen.  
You must visit the appropriate Nursecall Call Point to cancel the call. 

 

 

 
Deleting a Heat Detector from a

 

Touchsafe Pro 

Nursecall system

 

 

Select the heat detector ID to be deleted from the 

Delete Telecare Device

 pull-

down list box (prefixed with a “T

48

”).

 

Details of the heat detector selected will be displayed in the 

Room

House Code

 and 

ID

 boxes below. 

If the details shown are correct press the 

Delete

 

button.  

Press the

 

Close

 

button to exit. 

 
 

 

 
 

Refer to the Touchsafe Pro Display Panel user guide (Doc No. FM5505) provided with the Nursecall system for 
more information.  

 
 

 
 

Tynetec and Aidcall operate a policy of continuous improvement and reserve the right to change product specifications without notice. 

If any variation to the details contained in this document are suspected please contact Tynetec’s customer support dept on 01

670 352371.  

 

 

The 8 digit ID is printed on a label on the rear of the heat detector.  

Note: this ID will always start with 

48

xxxxxx.  

  

Touch the 

Add Telecare Device

 box then use the on-screen keypad to enter the ID.  

 

Hint: enter a zone in the 

Filter

 box to narrow the search for the room (see below).  

 

Select the users Call Point ID from the 

Room

 pull-down list box then press 

Save

Note:  this is the room unit that will be used to cancel the call. 

Use keypad to enter  

8 digit ID here 

Select the relevant  

Call Point ID here 

Save when done 

Close to exit mode 

On-screen keypad 

CALLING 
CANCEL 
HC 345  ID 001

 

Press the 

LIGHT GREY

  

button to cancel the call 

         

ZXT493

    W66794 

Ser: 10157951 
ID: 48046561 
EExxxx V1.00A 

MC 
 

 

Содержание Tynetec ZXT493

Страница 1: ...fixed to the ceiling or wall Use the fixing slots on the base plate as a guide when marking the position of the drill holes Fix with the screws and plugs provided Alternatively the base plate can be fixed with the self adhesive pads provided Ensure the surface is smooth clean and dry Press hard to make sure the self adhesive pads are properly adhered Position the detector on the base plate then tw...

Страница 2: ...possible within 30 days Sleep Easy Feature Low battery warnings often start at night or when it may be inconvenient to replace the detector You can silence the audible low battery chirp for a period of eight hours by pressing the Test button After this period the audible chirp will start again This process can be repeated a maximum of 10 times The detector will still detect heat during this time h...

Страница 3: ...e the unit will sound a Double Beep and the front light will return to STEADY GREEN The 8 digit ID can be found on a label on the back of the heat detector it will always start with 48xxxxxx The instructions given above for the standard Reach product also applies to the Reach Plus Reach Plus GSM and Reach Ultra models Refer to the instruction booklets provided with these products for more informat...

Страница 4: ...t then press the LINE key to clear the connection Refer to the Advent xt operator s manual Doc No FM0420 provided with the warden call system for more information on programming and testing radio devices Using the Manager s DECT telephone Press the LINE key and wait for the Ready prompt Enter 5 0 0 0 followed by the key Enter the flat number the heat detector is being installed in followed by the ...

Страница 5: ...alarm unit for more information on programming and testing radio devices Press and hold the TEST button the horn will sound and the light will flash RED The display will show Heat Alarm and the heat detector 8 digit ID number Check the ID is correct press the key to select Accept then press ENTER Note if another device has been picked up press ENTER to Reject and try again The display will show Fl...

Страница 6: ...yed in the Room House Code and ID boxes below If the details shown are correct press the Delete button Press the Close button to exit Refer to the Touchsafe Pro Display Panel user guide Doc No FM5505 provided with the Nursecall system for more information Tynetec and Aidcall operate a policy of continuous improvement and reserve the right to change product specifications without notice If any vari...