PP3 BATTERY
We declare this product complies with the relevant UK
legislation on condition that it is used in the manner
intended, and in accordance with the current installation
standards and/or the manufacturer's recommendations.
A copy of the full UKCA Declaration of Conformity is
available on request.
Mask Time & Battery Test
The Enuresis sensor transmitter has a 15 minute mask time to prevent the same alarm
call re-reporting several times. There is a facility for Engineers to override this mask
time and test the battery at the same time.
When the sensor is activated the alarm call will be transmitted once, no further
transmissions will occur for 15 minutes giving time for the cause to be cleared.
Engineer’s
mask reset function; open the case, press and hold
BUTTON 1
on the transmitter
for 2 seconds until the
RED
light flashes once. After a short delay the
RED
light will flash once more
to indicate the mask time is reset and the battery power is OK. If the
RED
light flashes several times the battery power is low.
Maintenance
The Enuresis sensor should be activated periodically to check the bed mat is functioning and verify operation.
Cleaning the Bed Mat
To prevent false alarms always un-plug the bed mat from the interface unit before cleaning.
The bed mat can be wiped clean with a damp cloth and soapy water or a mild disinfectant.
Do not fully submerge the bed mat lead or interface unit in water.
Replacement Bed Mats
Due to the nature of the product bed mats are non-returnable.
Replacement mats are available separately
–
order Tynetec Part No. W00369.
Battery Life
The battery is checked once a day, if the voltage falls and stays below a pre-set level for 7 consecutive days a low battery alert is
automatically transmitted. Once a low battery call is received the battery should be replaced within 7 days.
Detailed information on battery management for all Tynetec/Aidcall products is available - request Doc No. FM0630.
Battery Replacement
Slide the battery compartment cover off the rear
of the interface unit.
Disconnect the old battery and replace with an
identical type;
9V Duracell Plus MN1604
.
Refit the battery compartment cover.
Perform a test call to check the new battery is
fitted and working correctly.
Declaration of Conformity
SLIDE THE COVER OFF
TO REVEAL THE BATTERY
9 Volt
Battery
PRESS AND HOLD
BUTTON 1
TO RESET
THE MASK TIME
THE
RED
LIGHT
WILL FLASH
We declare this product complies with the relevant
European legislation on condition that it is used in the
manner intended, and in accordance with the current
installation standards and/or the manufacturer's
recommendations. A copy of the full CE Declaration of
Conformity is available on request.