Trouble shooting  

» The appliance does not work at all 

 
Check that the appliance is plugged in and that the power is on. 
 
Check that the fuse in the plug is intact. 
 

» The appliance is noisy 

 
Check that the appliance is level. 
 
Check that the appliance is not touching nearby items of furniture or other objects. 
 
“Whooshing” and “gurgling” noises caused by the refrigeration system are perfectly normal and cannot be 
avoided. 
 

»The appliance is not meeting the correct temperature 

 
Check that the door is shutting and sealing correctly. 
 
Check that the internal fan is working.

  

 
Check that the appliance is not too close to a heat source (such as a radiator). 
 
Check that the appliance is not overloaded with products.  The appliance requires space between stored 
items in order for the air to circulate correctly. 
 
Check that there has not been a sudden rise or fall in room temperature. The minimum and maximum 
ambient operating temperatures for your appliance are listed in the table on page 3 
 
Check that the appliance is positioned correctly. It should be positioned away from direct sunlight. There 
should be a minimum of 50mm clearance around the sides, 25mm at the top and 75mm of clearance at 
the back to allow free circulation of air. 
 
If the ideas given above have not solved the problem, call our service centre on 0844 8153755. Do not 
attempt repairs on your own. 

 

Disposal  

If you are disposing of an old cabinet, break off any latches and hinges, as a safe guard against small 
children trapping themselves inside. 
 
The refrigerant used in your appliance and the insulation materials require special disposal procedures in 
accordance with the current local regulations and laws. 
 
Ensure that none of the pipes on the back of the appliance are damaged prior to disposal. 
 
Details of the refrigerant can be found on the data plate. 
 

 

 

Содержание PGR1200

Страница 1: ...Page 1 MEDICAL REFRIGERATORS USER INSTRUCTIONS PSR600 PSR1200 PGR600 PGR1200...

Страница 2: ...Maintenance 7 Important Storage Advice Controller 8 Overview Button Functions 9 User Options Menu 10 User Settings Menu 11 Alarms Error Messages 13 Alarm History 13 Power Failure Other 14 Internal Lig...

Страница 3: ...0 3 C 0 3 C 0 3 C General Information Your appliance is suitable for the storage of pharmaceuticals and other material which requires a storage temperature between 2 C and 8 C The range of ambient tem...

Страница 4: ...position the front castors should be locked All models are supplied with fixing chains in order to anchor the units to the wall Fix the chains to the rear of the refrigerator using the bolts supplied...

Страница 5: ...switching on in order for the refrigerant to settle Once the appliance has been left to settle as above plug the unit into the mains and switch on at the plug The key switch on the rear of the applia...

Страница 6: ...ins plug Always pull the plug and never the cable The unit s cooling system contains refrigerant Both the unit and the products stored inside can be severely damaged if this system starts to leak Make...

Страница 7: ...the appliance with product Only store items on the shelves provided Store items tidily and allow for air to circulate around Make sure that stored items do not touch the reference bottle or the air se...

Страница 8: ...and to scroll through the menu options To select a menu option press To change a setting use the or buttons and confirm with To exit from the menu or return to the previous menu screen press the key T...

Страница 9: ...g displayed The controller will start the alarm tests SENSOR TESTING is displayed in the status line The controller will increase and decrease the load and air temperatures and the alarms will sound O...

Страница 10: ...COMPLETED is displayed Return to the main screen by pressing the MENU key until the temperatures are displayed again User Settings Menu The user settings menu gives access to settings which can be ch...

Страница 11: ...he appliance Alert alarms These are alarms which will require attention as soon as possible but do not affect the immediate safety of the stored items the refrigerator will continue to function as nor...

Страница 12: ...a fault with the load temperature display probe The refrigerator will continue to operate correctly but the load temperature display will be affected AIR SENSOR FAIL This message appears if there is...

Страница 13: ...checked EVENT OCCURRED will disappear from the status line on the controller Alarm events remain stored in the controllers memory for 7 days Use the UP and DOWN arrows to scroll through the list of a...

Страница 14: ...orescent tube Slide the new tube into the fixing slots figure 3 and secure by turning a 1 4 turn clockwise Clip the light cover back into place Plug in and restart the appliance Fig 1 Fig 2 Fig 3 Repl...

Страница 15: ...the appliance via a drain into the defrost water tray on the rear of the appliance this will evaporate and does not need emptying The appliance will display DEFROSTING when a defrost cycle is in progr...

Страница 16: ...port remove the plastic caps and the foam insert from the access port hole Slide the replacement plastic cap with a hole in over your probe as shown in photo 4A Place the cable of the probe in the sl...

Страница 17: ...at there has not been a sudden rise or fall in room temperature The minimum and maximum ambient operating temperatures for your appliance are listed in the table on page 3 Check that the appliance is...

Страница 18: ...tate the nature of the fault Please have as much information to hand as possible as this will help us in assisting you Contact Details Spares and After Sales Service E mail info gdha com Tel 0844 8153...

Отзывы: