Leander 950 Series: Скачать руководство пользователя страница 1

The 

950 

Series

Variable Height 

Multi-Purpose Motorized CPM Adjusting Table 

OPERATOR MANUAL

Mariendalsvej 22

8800 Viborg

Denmark

Tel. +45 8661 3611

[email protected]

www.chiroform.com

 

Содержание 950 Series:

Страница 1: ...The 950 Series Variable Height Multi Purpose Motorized CPM Adjusting Table OPERATOR MANUAL Mariendalsvej 22 8800 Viborg Denmark Tel 45 8661 3611 chiroform chiroform dk www chiroform com ...

Страница 2: ...ion in its uppermost position and the caudal pelvic section in its lowest position and at the lowest position of cam stroke DO NOT put the table in this position at any time while running the motor The table is not designed to use these to extreme flexion mode It is recommended that if you wish to increase the amount of flexion that you raise the center tube uplift variable height 2 3 inches with ...

Страница 3: ... Warranty 1 Warranty Procedure 1 Leander 950 12 Month New Product Limited Warranty 2 Warranty Terms Conditions 3 Installation 4 Space Considerations 4 Electrical Requirements 4 Cord Placement 5 Operation 6 Standard Features 6 Motorized Flexion Distraction 6 Isolated Flexion Drive System 6 Patient Flexion Off Switch 6 Lateral Flexion 7 Adjustable T bar 7 Hinged Head Section 7 Adjustable Face Cushio...

Страница 4: ...tioning 12 Check List 12 Maintenance Trouble Shooting 13 UPHOLSTERY 13 MECHLOKS 13 TABLE MAINTENANCE 13 Specifications 14 Weight 14 Length Width 14 Electrical 14 Miscellaneous 14 Shipping 14 Upholstery Limited Warranty Statement 15 FABRIC LEATHER ONE YEAR 15 Cushion Care Notice 15 DO NOT Use Cleaners Cleaning Wipes with Alcohol or Bleach Based Chemicals 15 ...

Страница 5: ...stom manufactured parts within 30 days after notification of the problem 5 The freight charges must be prepaid prior to shipment 6 An invoice will be generated for the cost of the product or part In most warranty situations the total cost will be credited with the exception of freight However there are occasions when the company will request that the defective part be returned to the factory In th...

Страница 6: ... or replacement of any part caused by negligent use or accidental damage or breakage 4 Repair or replacement of any part damage due to fire water or any other natural disaster 5 Labor to install or adjust non warranty parts 6 Labor to install any mechanical lock Mechlok after the first ninety 90 days If determined defective the Mechlok will be replaced free of charge and delivered from the factory...

Страница 7: ...r the cushion s Leander LLC will replace such cushion s free of charge Regardless of the cause all freight costs will be the responsibility of the customer 6 Term The new product limited warranty shall become effective upon receipt by Leander LLC of the shipping waybill acknowledging shipment 7 Charges a Leander LLC will not be responsible for reimbursing service personnel for any work that is not...

Страница 8: ...e 3 Placing your table in a room with carpet will reduce the noise caused by sympathetic vibrations that tend to be transmitted and amplified when the table is placed on a hard floor Also bear in mind that direct sunlight will decrease the life of and may discolor the upholstery over time Electrical Requirements The Leander 950 Table requires 110VAC 60Hz electrical power in the U S A This power co...

Страница 9: ...nnects via an international power entry module with a two pole switch for activating current see figure 4 Each of the control pedals attach via foot switch cord entry modules sockets see figure 4 the cord ends on the controls provided with the table are color coded to the appropriate connectors WARNING DO NOT PLUG POWER CORD INTO FOOT SWITCH SOCKET OR FUSE WILL BLOW Figure 4 Power Switch Power Cor...

Страница 10: ... table moves from horizontal to horizontal in 2 seconds completing 30 cycles each minute A table equipped with an EVS may have the cycling frequency adjusted Isolated Flexion Drive System The Isolated Flexion Drive System smoothens out the transfer of power from the gearbox to the caudal section and reduces noise and vibration normally associated with flexion distraction tables It includes vibrati...

Страница 11: ...atients are able to vary the amount of distraction force applied to the spine by pulling against the T bar with lighter or heavier force Under the front edge of the right head cushion is a chrome lever which controls the T bar locking mechanism see figures 6 9 To extend the T bar press the lever toward the rear of the table With the lever depressed the T bar may be placed anywhere in the available...

Страница 12: ...uld be in the lowered position To lower the abdominal section support the abdominal cushion with one hand and lift the abdominal release lever see figure 12 Gently lower the cushion Important without properly supporting the cushion patient patient discomfort could result Adjustable Ankle Support At the foot of the table is an adjustable ankle cushion that may be extended up to 10 inches to accommo...

Страница 13: ...ft features of the table are both operated by foot switches The flexion motor is started or stopped by simply pressing the Start or Stop side of the On Off pedal The end of the pedal away from the cord is the Start side and is marked with a green dot The end of the pedal at the cord is the Stop side and is marked with a red dot The uplift adjusts to any preferred height by pressing and holding the...

Страница 14: ...ront section remains down see figure 19 To raise or lower the caudal section support the weight of it with one hand and with the other hand squeeze the control located under the left side similar to the adjustable ankle support Position the section as desired and release the control to lock it in place Important Figures 18 19 demonstrate the positioning combination extremes available Improper posi...

Страница 15: ... of the table on the upper portion of the base see figure 17 It allows the therapist to pre determine exact treatment times for each client and easily repeat that treatment at each visit A manual override is built in to allow operation with or without automatic stopping To use the flexion timer 1 Turn the timer knob to the desired setting Remember to turn past 4 minutes and back to set for 1 to 3 ...

Страница 16: ...l section could injure your patient 4 Adjust the ankle cushion so that it just touches the top of the instep of the foot Patients may then increase the amount of distraction force applied to their spine by pulling against the ankle cushion with their toes The ankle straps may be used to secure the patient s legs and in certain cases may increase patient traction 5 Extend the T bar until the patien...

Страница 17: ...here may be some issues that may occur The most common is motor sounds This product contains a Leeson Motor We have found it to be the best quality and most durable Motor capacity is approximately 350 pounds Like all electrical products they are subject to various currents that can cause the motor to make a louder sound at times If you feel the motor is increasing in sound within the first 6 month...

Страница 18: ...nal voltage Miscellaneous Flexion Distraction Travel 12 2 degrees or 6 3 4 inches measured at the end of the retracted ankle cushion Flexion Distraction Speed 0 30 cycles per minute Lateral Flexion Travel 25 degrees to each side Uplift 18 1 2 to 24 inches in 11 seconds or 21 29 inches in 16 seconds Gearbox Lubrication Mobile SHC 634 Synthetic Gear Reducer Lube Upholstery Yardage 2 3 yards total Mi...

Страница 19: ...natural markings shrinkage stretching or variations in grain or color which occur naturally and prove the authenticity of the leather Damaged cushions will be sent to fabric manufacturing lab for alcohol chemical testing prior to approval of warranty claim approval or service CUSHION CARE NOTICE DO NOT Use Cleaners Cleaning Wipes with Alcohol or Bleach Based Chemicals Recommended cleaner 40 Liquid...