5. HOW TO USE
19
1. File
Open
Opens a setup file that has been saved.
Save
Saves the settings with a name.
2. Window No.1 to 32
EN
Select the check box to enable the measurement range and allow it to be selected with
Window No. in the measurement window.
Start Line / End Line
Enter the measurement start and end lines in the range of 21 to 263 and 283 to 525,
respectively.
Enter a larger number for End Line than Start Line. Also, set the difference between
the start and end lines to 20 or less.
Start Time / End Time [us]
Enter the measurement start and end times in the range of 5.0 to 63.5 assuming the
rising edge of the horizontal sync signal to be 0 [us].
Enter a larger number for End Time than Start Time.
SET This Window
Apply the settings for each window.
3. INIT COLOR BAR
Sets Window No. 1 to 8 to the measurement ranges appropriate for the color bar.
To apply the settings, click SET ALL.
4. SET
ALL
Applies the settings of all windows.