LDetek LDP1000 Скачать руководство пользователя страница 11

 

 

LDP1000 User’s manual

 

 

 

10

Step 5 :  

 

 
Step 6 : 

 

 
Step 7 : Install the new getter by re installing the screws of step 6. 
 
Step 8  :Put back the wires in the terminal bloc J1 and J2 that you have removed in step 5. 
Use wiring diagram as reference for the cable #. 
 
Step 9 : Re install front part of the LDP1000 and put back the screws that have been 
removed in step 1. 
 
Step 10 : You can put back the gas lines in and out and power up the unit for normal 
operation running. 
NOTE : MAKE SURE TO WELL PURGED THE LINE BEFORE TO CONNECT TO 
THE INLET (10 MINUTES PURGE). DO NOT EXPOSE THE DEVICE TO AIR; 
MAKE SURE TO QUICKLY CONNECT INLET AND OUTLET TO AVOID AIR 
CONTAMINATION OF THE UNIT. IF THIS NOTE ISN’T PREPERLY RESPECTED, 
SOME DAMAGE CAN OCCUR TO THE LDP1000 DEVICE. 
FOR THE HIGH PURITY VERSION, A BRAND NEW GASKET MUST BE 
INSTALLED ON THE INLET AND OUTLET FITTINGS. 

Using a screw driver, 
remove the wires in J1 
and J2 terminal blocks. 

Using a srew driver, remove 
the getter isolated box by 
removing the 4 screws on 
the back side plate. 

Содержание LDP1000

Страница 1: ...LDP1000 Gas Purifier for noble gases and nitrogen USER S MANUAL V1 0 Printed in Canada Copyright 2010...

Страница 2: ...ent 1 Forwarning 2 2 Warranty maintenance and service policies 3 3 Specifications 6 4 Cautions and installation information 6 4 1 Gas line connection 6 4 2 Installation steps 6 5 Lifetime and LED indi...

Страница 3: ...tek makes assumption that all operators have taken the time to read this information prior to install operate and troubleshoot the purifier If any error is suspected by the reader please contact LDete...

Страница 4: ...r the loss or damage directly or indirectly arising from the use of the equipment of goods from breach of any warranty or from any other cause All other warranties expressed or implied are hereby excl...

Страница 5: ...VERED BY UMBRELLA INSURANCE AND OR INDEMNITY COVERAGE PROVIDED TO BUYER ITS ASSIGNS AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HERE UNDER Major force Seller is not liable for failure to perform...

Страница 6: ...tions may apply to international shipments PROPRIETARY RIGHTS Buyer agrees that any LDetek s software firmware and hardware products ordered or included in the goods ordered are proprietary of LDetek...

Страница 7: ...tion Maximum 200 watts Dimensions 12 304 8 high 4 12 104 mm deep 4 25 108 mm wide Weight 5 lbs 2 26 kg 4 Cautions and installation information 4 1 Gas line connection Be sure to connect the upcoming g...

Страница 8: ...make sure you have the right operating voltage 7 With the power on the purifier let the gas vent to the atmosphere for about 3 hours 5 Lifetime and LED indication The LDP1000 getter has a lifetime of...

Страница 9: ...nual 8 6 Drawings Wiring Diagram Cable 4 Cable 5 Cable 2 Cable 3 RS 232 Connector Cable 1 Thermocouple 1 Thermocouple 2 Yellow Led Green Led Red Led Heater 1 Heater 2 Pin A Corcom Pin C Module Pin B P...

Страница 10: ...he lifetime is over Getter replacement procedure Step 1 Turn off the power on the LDP1000 unit Step 2 Remove the inlet and outlet tubing connections Step 3 Step 4 Remove the front part of the LDP1000...

Страница 11: ...power up the unit for normal operation running NOTE MAKE SURE TO WELL PURGED THE LINE BEFORE TO CONNECT TO THE INLET 10 MINUTES PURGE DO NOT EXPOSE THE DEVICE TO AIR MAKE SURE TO QUICKLY CONNECT INLET...

Страница 12: ...LDP1000 User s manual 11...