Step 5 :
Step 6 :
Step 7 : Install the new getter by re installing the screws of step 6.
Step 8 :Put back the wires in the terminal bloc J1 and J2 that you have removed in step 5.
Use wiring diagram as reference for the cable #.
Step 9 : Re install front part of the LDP1000 and put back the screws that have been
removed in step 1.
Step 10 : You can put back the gas lines in and out and power up the unit for normal
operation running.
NOTE : MAKE SURE TO WELL PURGED THE LINE BEFORE TO CONNECT TO
THE INLET (10 MINUTES PURGE). DO NOT EXPOSE THE DEVICE TO AIR;
MAKE SURE TO QUICKLY CONNECT INLET AND OUTLET TO AVOID AIR
CONTAMINATION OF THE UNIT. IF THIS NOTE ISN’T PREPERLY RESPECTED,
SOME DAMAGE CAN OCCUR TO THE LDP1000 DEVICE.
FOR THE HIGH PURITY VERSION, A BRAND NEW GASKET MUST BE
INSTALLED ON THE INLET AND OUTLET FITTINGS.
Using a screw driver,
remove the wires in J1
and J2 terminal blocks.
Using a srew driver, remove
the getter isolated box by
removing the 4 screws on
the back side plate.