![lbx H03D Скачать руководство пользователя страница 11](http://html.mh-extra.com/html/lbx/h03d/h03d_user-manual_681698011.webp)
11
6.2 Residual heat warning (HOT)
In order to prevent burns from the hotplate, the digital hotplate has a residual heat warning
function. When the heating function is switched off and the heating plate temperature is still above
50°C, “Hot” will flash on the LED screen to warn that there is a hazard of burns from the hotplate.
When the unit is powered off, the LED screen alternates the temperature of the hotplate and “Hot”.
When the hotplate temperature drops to below 50°C, the unit will automatically switch off. If users
want to turn off the LED immediately, just pull out the plug directly. When the plug is pulled out, the
residual heat warning function stops running.
7.
Function: Stirring
The function “stirring” of LED digital model (H03D and S03D):
•
Turn on the device by pressing the switch ON/OFF.
•
Start the function "stirring" by pressing the "STIR" knob and adjust the speed slowly until the set
value.
•
When starting the function "stirring", the display shows set value and then blinks. This blinking
will stop only when the actual speed reaches the set value.
•
At the end of the manipulation, turn the knob to the left to slow down and turn off "stirring" by
pressing the knob.
•
Turn off the device by pressing the switch ON/OFF.
The function “stirring” of classic model (S03):
•
Turn on the device by pressing the switch ON/OFF.
•
Activate and adjust the speed of stirring by slowly turning the speed control to the set value.
•
At the end of the manipulation, turn the knob fully counterclockwise to turn off the stirring.
•
Turn off the device by pressing the switch ON/OFF.
8.
Faults
•
Instruments cannot be power ON.
Check whether the power line is unplugged.
•
Fault in power on self-test
Switch OFF the unit and then switch back ON.
•
Stir speed cannot reach set point
Excessive medium viscosity may cause abnormal speed reduction of the motor
•
Unit cannot be powered off by pressing the switch on / off.
Check if the residual heat warning function is still ON and hotplate temperature is above
50ºC (the LED screen still works and “Hot” appears on the screen).
If these faults are not resolved, please contact manufacturer/supplier.
Содержание H03D