The setting will exit automatically after 5S without
operation
36.8
Leave the bath to stabilize and check the calibration
High temperature point calibration:
For example:
Before calibration, the screen displays 60°C but the actual liquid temperature is
65.0°C;
After calibration, the screen displays 64.7°C and the actual liquid temperature is
65.0°C;
Operation
Display
°C
Select the menu option
HCP
Select the high temperature point calibration
setting menu option
60.0
Press set to enter the calibration value
65.0
Enter the reference thermometer value
SET
Press set to store the calibration value
The setting will exit automatically after 5S without
operation
64.7
Leave the bath to stabilize and check the calibration