Lawson 17B Скачать руководство пользователя страница 8

 5.  Be sure the G terminal on the Model 17B is connected to the
     computer chassis ground or an appropriate alternate ground.

D.  If channel 0 of the multiplexer is always read regardless of
    which channel of the Model 17B is selected by the output
    control lines, check to be sure that the Guard pin on the A/D
    card is connected to the G terminal on the Model 17B.

SECTION 4.    SPECIFICATIONS

SIGNAL INPUTS:          16 channels
                        Both plus and minus inputs
                        are switched

SIGNAL INPUT RANGE:     +/- 5.5 volts

SWITCHING TIME:         0.5 ms maximum

SWITCH RESISTANCE
TYPICAL:                On resistance 20 K ohms
                        Off resistance 100 Megohms

INPUT PROTECTION:       All inputs protected against
                        overvoltages of up to 130 volts
                        transitory or 60 volts continuous

CONTROL INPUTS:         Fully compatible with all
                        Lawson Labs A/D cards

TYPICAL CONTROL
INPUT CURRENT:          0.06 milliamps
                        Active low

POWER REQUIREMENT:      115 VAC, 6 Watts

SIZE:                   6.25 x 3.75 x 2 inches

                     LIMITED WARRANTY

     The Lawson Labs Inc. Model 17B Differential Multiplexer is
guaranteed against defects in materials and workmanship for a
period of one year from the date of delivery. Products must be
returned to Lawson Labs for warranty service. Contact Lawson Labs
at 800 321-5355 for return authorization before returning
anything for service.

     The above warranty is in lieu of all warranties express or
implied. Lawson Labs, Inc. will not be liable for indirect or

Содержание 17B

Страница 1: ...uy your excess underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing R...

Страница 2: ...tion with Model 66 Expansion Modules up to 32 multiplexers can be handled by a single A D card The control output lines on the A D determine which input is selected Both the plus and minus connections...

Страница 3: ...her input is used the control codes should be adjusted as described in the A D manual The ground terminal G on the Model 17B should be connected to chassis ground at the computer WARNING Use a cover s...

Страница 4: ...d the A D will be overranged Connect the plus and minus wires from the first input channel to be switched to the input terminals labeled 0 and 0 on the Model 17B Before attaching additional inputs you...

Страница 5: ...g the A D channel selection code plus the number of the desired multiplexer channel as the control code For example with the multiplexer at A D channel CH if you are using a Model 134 or 140 the instr...

Страница 6: ...0 IF X AND 64 THEN V V 4070 V V 25 4080 RETURN Include this subroutine if you are using a Model 134 In line 4030 you may set AVERAGE anywhere from 1 to 255 Larger numbers reduce noise while smaller nu...

Страница 7: ...you are using a wall socket ground for the external ground connection make sure that it is actually at ground potential 5 If more than one multiplexer thermometer or other device is connected to the...

Страница 8: ...e 100 Megohms INPUT PROTECTION All inputs protected against overvoltages of up to 130 volts transitory or 60 volts continuous CONTROL INPUTS Fully compatible with all Lawson Labs A D cards TYPICAL CON...

Страница 9: ...consequential damages caused by any defect in this product Some states do not allow the limitation of consequential damages so this exclusion may not apply to you...

Страница 10: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Отзывы: