background image

M•AD-824  

           

      Quick-Start Guide 

 

 

 

 

 

5   

 

8.) Select DITHER & Optional NOISE SHAPING for 16 bit or 20 bit modes. 

 
Dither, ABC-1 and ABC-2 are grouped together (similar to word length). 

αβC

αβC

αβC

αβC α

coustic 

β

it Correction®

 

is dither plus psycho-acoustic enhancement of the 

converter’s dynamic range for word length reduction to 20 or 16 bits. Some describe 
this as getting more than 16 bits performance with a 16 bit word length. 

 ABC-1

 provides a gentler enhancement curve. 

 ABC-2

 provides a more aggressive enhancement curve. 

 

•It is possible to set this function to 

none

 of the three choices for “no dither or 

αβC

αβC

αβC

αβC

.” In this 

case, all three lamps will be blinking. 
 
•Setting the position of the internal DIP switch position to “ON” changes ABC-1 and ABC-2 
to less aggressive curves. Please refer to the LavryBlue Operations Manual.

 

 
9.) METER FUNCTIONS 
 
To Use The Reference Meter Bridge: 

Please Note: To use this mode, you need a reference tone source (for example 1kHz) at 
“0dB” on your VU meters or at a level that represents the AVERAGE level of the music 
source. This level will typically be in the range of 10 to 20dB’s below analog PEAK level. 
The range of the meter is “expanded” around the reference level so that each lamp above or 
below the reference level indicates approximately 0.2dB of level change. 
 
Select “Reference.” A pair of solidly lit lamps (L and R) indicates the reference level 
between -10 and -20dB on the Meters. Click up (Set) to step to the desired REFERENCE 
LEVEL (-10, -12, -14, -16, -18 or –20dBFS). Apply the tone to the analog inputs of the 
M•AD-824. If the top lamps of the meter are blinking, it means that the analog input level is 
ABOVE the meter range, and if the bottom lamps are blinking the tone is BELOW the range. 
Use a small screwdriver to adjust the front panel 20-turn pots until the second pair of lamps 
stops blinking, and move up or down the scale to coincide with the selected reference level. 

 

Determining the optimum level setting for the trimmers of the MAD module will take some 
experimentation. The most important consideration is that the typical peak level of the audio 
program you want to record comes very close to, but does not exceed “0dBFS” which is 
represented by the “MAX” lamps on the MAD meters. The following procedure uses music 
program to roughly determine the optimum level; then allows you to “fine tune” it using the 
REFERENCE function.  

 

Once you have established the output gain of the device feeding the MAD module, DO NOT 
change it during this calibration procedure. 
 

1.)

 

Feed music program to the analog inputs which is typical of the type of material you 
want to record with the MAD-824. Adjust the trimmers on the front of the MAD 
module so that the peak level is close to “0” on the MAD meters, but does not cause 
the MAX lamps to illuminate. 

Содержание M AD-824

Страница 1: ...Model M AD 824 Analog to Digital Converter and M SYNC Quick Start Guide Version 2 0 ...

Страница 2: ...UT CABLES to the M AD 824 Looking at the back of the 4496 each M AD 824 module has a group of three XLR connectors In each group the LEFT INPUT is the right female XLR and the RIGHT INPUT is the left female XLR connector The male XLR is the AES digital output This mirror image position of the analog inputs is due to the straight through construction of the modules Facing the front of the 4496 the ...

Страница 3: ...indicated when the lock lamp is illuminated AES to lock the M SYNC to an incoming external AES signal The lock condition is indicated when the lock lamp is illuminated PLEASE NOTE The M SYNC s BNC Sync input is NOT internally terminated For a simple two device system connected with a short BNC cable termination is probably not necessary For example connecting two 4496 units together Sync Out to Sy...

Страница 4: ...4 The SET SELECT switch is intuitive and actually far easier to use than to describe Use it to enter or exit the Program Mode by holding the switch either up or down for 3 seconds There are individual functions Saturation Modes Reference Peak Hold and multiple parameter functions Wordlength Dither ABC SELECT click switch Down to moves the indicator from one function to the next In all cases except...

Страница 5: ...on that emulates analog tape saturation is applied to the waveform As a result signal peaks that would have reached 0dBFS are reduced by 6dB s Please keep in mind that enabling this mode automatically adds 6dB more level as seen on the meters of the M AD 824 for signals below the 12dBFS threshold so the point at which the saturation starts is 6dB on the meters The MAX lights still indicate when th...

Страница 6: ...el change Select Reference A pair of solidly lit lamps L and R indicates the reference level between 10 and 20dB on the Meters Click up Set to step to the desired REFERENCE LEVEL 10 12 14 16 18 or 20dBFS Apply the tone to the analog inputs of the M AD 824 If the top lamps of the meter are blinking it means that the analog input level is ABOVE the meter range and if the bottom lamps are blinking th...

Страница 7: ...example where the reference mark is at 12 this level would be indicated by the lamps that are between the 30 and 20 label on the MAD meters which is five lamps below the 12 lamp 6 Because the level established using music program in step 1 above represent the loudest level without clipping it would be advisable to use the lower of the two channels as the reference level For example before further ...

Страница 8: ...e Slave modules differently than the Master be sure to program the Master module first When the frame is powered the slaves will always emulate the nearest master module INSTALLATION OF THE M DA 824 The flexibility of the LavryBlue 4496 chassis allows up to 4 single width modules to be in different slots within the frame with some restrictions If the unit has an M AD 824 A to D module it must be i...

Отзывы: