Maintenance and Decontamination
No specific maintenance is required except for periodic check of calibration.
Cleaning and decontamination may be done with decontamination solutions suitable
for use with ABS polymer and polycarbonate materials.
Replacement of the alkaline battery
Open the battery compartment cover with a coin or small screwdriver.
Remove the old battery and replace with new one.
The + pole goes in first.
Use only AAA alkaline battery (IEC LR03).
Close the cover.
After battery replacement the dosimeter automatically turns on.
Note!
After fitting a fresh battery it is recommended to perform the functional
test (dIA). If the battery symbol appears on the display after the test, refit the battery.
Radiation detected:
Gamma and X-rays
Measurement range:
dose 1 µSv - 9.99 Sv
0.1 mrem - 999 rem
dose rate 5 µSv/h - 3 Sv/h
0.5 mrem/h - 300 rem/h
Measurement accuracy:
better than ±5% (Cs-137, 662 keV)
Energy response:
60 keV - 3 MeV, better than ±25%
Dose rate linearity:
better than ±15%
Operational temperature range:
-20°C to +50°C
Operational humidity:
10 - 90% RH
Dimensions:
78 * 67 * 22 mm, excluding pocket clip.
Weight:
80g, including battery
Specifications