OX-TRAN Model 2/22 Operator’s Manual
Testing Flat Films
MOCON, Inc.
Revision R
5-7
Loading and Unloading the Test Cell
The OX-TRAN Model 2/22 uses a pneumatic system to clamp, unclamp and eject the Test Cell. The Test
cell must be manually loaded and unloaded from the instrument.
To Unload the Test Cell from the Cell Tray follow the instructions below (refer to Figure 5-3):
1.
Press the Cell Load/Unload button on the front of the instrument (see Figure 2-1).
2.
Open the Cell Tray completely by gently pulling straight back from the front panel.
3.
Grasp the Test Cell by the front and back edges and lift straight up.
To Load the Test Cell into the Cell tray follow the instructions below (refer to Figure 5-3):
1.
Grasp the Test Cell by the front and back edges and lower it straight down.
2.
Close the Cell Tray completely by gently pushing straight towards the front panel.
3.
Gently hold the Cell Tray Shut with a finger.
4.
Press the Cell Load/Unload button to clamp the cell (see Figure 2-1).
Figure 5-3: Loading the Test Cell