86 / 134
ECO Gold
The adjustments described in this appendix are only intended for specially qualified personnel.
13
Other settings
13.1
Resetting to factory settings
-
Access to the main menu level is obtained by pressing the enter
key .
-
Selection and confirmation of
Setup
Factory Setting
.
The adjacent menu window appears.
-
Select the menu item
all default
.
Select one of the following options:
-
Selecting
no
returns to the previous display without making any
changes.
-
Selecting
yes
restores the factory settings if you confirm this
with the enter key.
-
By selecting
Control
you can select the displayed parameters
with or .
Select the appropriate menu item in the parameter list.
-
The internal and the external control parameters can be reset
using
Control parameter
.
-
The settings for the internal sensor can be reset with
internal Pt100
.
-
With
miscellaneous
the following can be reset: set value, pump
level, max. current consumption, control to internal and auto-
start to "auto".
Select one of the following options:
-
Selecting
no
returns to the previous display without making any
changes.
-
Selecting
yes
resets the selected parameter if you confirm this
with the enter key.
-
By pressing or
(ESC) you are returned to the menu level without any change.
Содержание Viscotemp 15 G
Страница 77: ...11 Declaration of Conformity 09 2021 ECO Gold 77 134 ...
Страница 78: ...78 134 ECO Gold 09 2021 ...
Страница 81: ...09 2021 ECO Gold 81 134 W Warnings 53 92 X Xp 106 Xpf 108 ...
Страница 82: ...82 134 ECO Gold 09 2021 ...
Страница 83: ...09 2021 ECO Gold 83 134 ...
Страница 84: ...84 134 ECO Gold 09 2021 ...
Страница 85: ...09 2021 ECO Gold 85 134 Appendix with settings ...
Страница 121: ...09 2021 ECO Gold 121 134 Click on the tab with the mouse and then on the Device manager ...
Страница 129: ...09 2021 ECO Gold 129 134 ...
Страница 130: ...130 134 ECO Gold 09 2021 ...
Страница 131: ......