Personnel:
n
Operating personnel
1.
Change to the main menu.
2.
Select the menu item
Setup
Basic setup
Autostart
.
3.
Select one of the following options:
n
With
off
, the device is in the standby operating mode
after a mains power failure.
n
Normal operation is continued directly after a power
failure using
on
.
4.
Confirm with the OK button.
6.12.5
Limiting current consumption
If your line fuse is below 16 A, the current consumption can be
gradually reduced from 16 A to 8 A. For devices with a heater, the
output of the heater is reduced accordingly. Take account here
whether other consumers are still connected to the same fused cir-
cuit or whether your device is the only consumer.
Personnel:
n
Operating personnel
1.
Change to the main menu.
2.
Select the menu item
Settings
Basic settings
Curr. Consumpt.
3.
Adjust the current consumption accordingly.
4.
Confirm with the Enter button.
Fig. 45: Specifying Autostart
Fig. 46: Specifying current consump-
tion
Operation
Variocool
65
Содержание VC 1200
Страница 47: ...Fig 13 Menu structure part 1 Operation Variocool 47...
Страница 48: ...Fig 14 Menu structure part 2 Menu structure for Graph Clock and Standby Operation Variocool 48...
Страница 103: ...General Variocool 103...
Страница 110: ......
Страница 111: ......