1 Quick Installation Guide
The LAUDA.LIVE wireless gateway enables access to the LAUDA Cloud
independently of the customer network LAUDA.LIVE and uses a cellular
connection for this purpose. The gateway is connected to the LAUDA
device via the Ethernet interface.
The gateway is preconfigured for LAUDA Ultracool devices.
Fig. 1: Principle of the cloud connection
1.
Connect the supplied antenna to the antenna socket of the gateway.
2.
Connect the gateway to the LAUDA device via the Ethernet inter-
face. You can use the supplied Ethernet cable (1.5 m) for this purpose.
If the cable is too short, you can use any standard CAT5.E LAN cable
to connect the gateway to the LAUDA device.
3.
Supply the gateway with voltage via the supplied power unit. To do so,
connect the power supply unit to the power socket of the gateway.
4.
The Ethernet IP address and the subnet mask of the LAUDA Ultra-
cool device must be configured to static, so that the wireless gateway
can access LAUDA.LIVE (please refer to the Ultracool operating
instructions for further information):
IP: 172.20.12.155
Subnet mask: 255.255.0.0
The gateway is preconfigured to the IP address
172.20.12.1. If the described IP addresses cannot be
used, due to customer-specific requirements, please
contact LAUDA Service. This particularly applies if the
LAUDA devices are to be integrated LAUDA.LIVE
locally in the network (e.g. in addition to a control
station computer or web server).
Ultracool 4G Gateway
Installation
Fig. 2: Gateway with antenna
Fig. 3: Gateway with power supply
1
0
/
24
Ultracool 4G Gateway and M2M SIM cards
V1