ALARMS AND WARNINGS LIST
See the following table for the main possible causes for an alarm or warning and their solution:
FAULT
CAUSE
SOLUTION
RESTART PROCEDURE
AL145, 158, 462 or
546
HP alarm by pressure
switch
Alarm due to high
pressure of the
refrigerant:
T
he pressure
of the fridge circuit is
higher than the
maximum allowed. It
stops the compressor
Panel(s) of the
housing open (only
UC 8 to UC 100)
Low airflow into the
condenser
The ambient
temperature is too
high
Motor fan not
working
High pressure switch
failure
Close the panels
Check that there is enough free
space in front of the condenser
and clean the condenser if
necessary
Wait until the ambient
temperature is lower
The motor fan is not working if it is
not turning when the chiller is
running and then tripping by this
alarm. Contact authorized
technical service
Contact authorized technical
service
Disconnect the chiller by
turning Off the main power
switch (see point 5.1). Turn it
back On after at least 30
seconds have passed
AL165
Alarm freeze evaporating
temp.
Alarm due to low
evaporating temperature:
The evaporation
temperature is too low
and there is risk of
freezing. It stops the
compressor if it happens
3 times in a short period
Water circuit blocked
Possible freezing due
to low ambient
temperature
Water tank
temperature sensor
fault
Clean the water circuit. If
necessary replace the water filter
element. Check for closed valves
in the circuit
The ethylene glycol concentration
must be according to point 4.1
and the antifreeze setpoint also
has to be adjusted accordingly.
Contact authorized technical
service
Measure the water temperature
inside the tank and check that it is
approximately the same as shown
on the controller
’s display. If it isn’t
contact authorized technical
service
Disconnect the chiller by
turning Off the main power
switch (see point 5.1). Turn it
back On after at least 30
seconds have passed