![Lauda RE 1050 G Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/lauda/re-1050-g/re-1050-g_operating-instructions-manual_4088318036.webp)
ECO GOLD
36
Putting the device into operation
28.04.10/ YACE0088
7.4
Basic setup
7.4.1
Setting the overtemperature switch-off point T
max
−
Press the key
and keep it pressed during the com-
plete setup; simultaneously press
. The entry window
appears. The T
max
value is underlined. The maximum and
minimum possible adjustable temperature values are dis-
played.
−
Change the value with
or
.
−
By
pressing or
individual numbers can be selected
and changed with
or
.
−
The selected value is confirmed with
.
−
By
pressing
you are returned to the menu level
without the value being changed.
7.4.2
Setting the temperature setpoint value
−
Access to the main menu level is obtained by pressing
.
−
"Setpoint Value" is highlighted in colour.
- - -
סּ
OK
- - -
Tmax
Max. Temperature
Max:
205
Min:
0
85
ESC
סּ
OK
Standby
Main Menu
T
SET
Setpoint Value
Settings
►
Programmer
►
Interfaces
►
Graph
►
Clock
►
Standby