Lauda Puridest PD 2 D Скачать руководство пользователя страница 11

09/2020

Puridest Bi Distillation Apparatus

11 / 32

Cooling water outlet

7.4

Connect a ¾" temperature-resistant hose to the cooling water outlet. It must not be longer than approx. 1.5 m. The hose
must be led into a drain on a lower level and must have a slope on its complete length. It must be ensured that the cooling
water flows off without causing backwater.

Caution:
The cooling water leaves the Distillation Apparatus with temperatures of up to 70 °C.
Danger of scaldings!

Drain Bi-stage boiler

7.5

A ½" hose to drain the boiler for cleaning or maintenance can be connected.

Caution:
The water leaves the boiler of the Distillation Apparatus with temperatures of up to 100 °C.
Danger of scaldings!

Drain Mono stage boiler

7.6

A ½" hose to drain the boiler for cleaning or maintenance can be connected.

Caution:
The water leaves the boiler of the Distillation Apparatus with temperatures of up to 100 °C.
Danger of scaldings!

Water connections pos.

 and

 are parts of an accessory. They are only fitted if a separate water

supply was specifically ordered with the unit. The depicted supply valve pos.

 is only required if the

unit is equipped with a separate water supply.

8

Initiation

After assembly of all power and water connections, and to initiate the Distillation Apparatus, the stop valves of the water
supply (

 and

) must be opened and the main switch (I) must be switched on. The first litres of produced distillate

(mono and bi distillate) should not be used. Between switching the Distillation Apparatus off and then on again, always wait
for approx. 10 seconds as the unit will otherwise be switched into "Clean" mode.
During down-times of the Distillation Apparatus, all cut-off valves of the water supply should always be closed.

Содержание Puridest PD 2 D

Страница 1: ...Operating Instructions Puridest Bi Distillation Apparatus PD 2 D PD 4 D PD 8 D...

Страница 2: ...2 32 Puridest Bi Distillation Apparatus 09 2020...

Страница 3: ...in switch II Pilot lamp Operation Mono stage III Pilot lamp Low water Mono stage IV Pilot lamp Clean V Pilot lamp Operation Bi stage VI Pilot lamp Low water Bi stage Before installation please check w...

Страница 4: ...4 32 Puridest Bi Distillation Apparatus 09 2020...

Страница 5: ...water supply 10 7 3 Cooling water outlet 11 7 4 Drain Bi stage boiler 11 7 5 Drain Mono stage boiler 11 7 6 8 Initiation 11 9 Functional Description 12 Distillation Apparatus for double distillation 1...

Страница 6: ...the mains supply examples 22 14 1 14 1 1 PD 2 D for 230 V 22 14 1 2 PD 4 D and PD 8 D for 220 V 3 at Mains supply 230 V 3 PE 50 60 Hz 23 14 1 3 PD 4 D and PD 8 D for 220 V 3 at Mains supply 230 V 3 PE...

Страница 7: ...led water leaves the Distillation Apparatus with more than 90 C Danger of scaldings Depending on the tap water quality the produced mono distillate has a conductivity of approx 2 2 S cm at 25 C The pr...

Страница 8: ...Distillation Apparatus PD 2 D net weight 20 4 kg PD 4 D net weight 27 5 kg and PD 8 D net weight 43 0 kg must be lifted carried and transported to the location by at least two persons The Distillation...

Страница 9: ...d back onto the unit and secure with the two screws and four quick release catches Then install the glass fittings Screw together the glass valve with the mono distillate outlet left hand front of the...

Страница 10: ...n accessory Only fitted if a separate water supply was ordered The water inlet for the separate water supply feeds pretreated water to the boiler of the Distillation Appa ratus through a solenoid valv...

Страница 11: ...ing or maintenance can be connected Caution The water leaves the boiler of the Distillation Apparatus with temperatures of up to 100 C Danger of scaldings Water connections pos and are parts of an acc...

Страница 12: ...valve for withdrawal of mono distillate is closed the mono distilled water is led into the boiler of the bi stage Once the heating elements in the bi stage are below water level the heating elements...

Страница 13: ...tage boiler and rinse the boiler with clear water several times Commercial descaling agents suitable for use with stainless steel can also be used according to the manufacturer s in structions e g rea...

Страница 14: ...f the water in the boiler caused by the distillation process in the mono stage boiler foam will form during evaporation of water Once the foam gets in contact with the electrode in the mono stage boil...

Страница 15: ...ee of charge and will have it materially recycled Before the unit is returned a legally binding declaration must be provided from the sender confirm ing that the unit is free from harmful and or hazar...

Страница 16: ...C 15 K Electromechanical temperature limiter with capillary tube sensor Switch off temperature 135 C 15 K Water pressure min max 3 bar 7 bar 43 5 psi 101 5 psi 3 bar 7 bar 43 5 psi 101 5 psi Electrica...

Страница 17: ...ctromechanical temperature limiter with capillary tube sensor Switch off temperature 135 C 15 K Water pressure min max 3 bar 7 bar 43 5 psi 101 5 psi Electrical connection Mains supply On site fuse 22...

Страница 18: ...amp Clean LED 6 V DC H2 Pilot lamp Low Water mono stage LED 230 V AC H3 Pilot lamp Operation mono stage LED 230 V AC H4 Pilot lamp Low Water bi stage LED 230 V AC H5 Pilot lamp Operation bi stage LED...

Страница 19: ...09 2020 Puridest Bi Distillation Apparatus 19 32 PD 4 D version 220 V 3 13 2 PD 4 D version 400 V 3 13 3...

Страница 20: ...20 32 Puridest Bi Distillation Apparatus 09 2020 PD 8 D version 220 V 3 13 4 PD 8 D version 400 V 3 13 5...

Страница 21: ...PD 4 D PD 8 D Mains supply 220 V 3 N PE 50 60 Hz Model PD 4 D PD 8 D Mains supply 400 V 3 N PE 50 60 Hz ge gr yellow green PE Protective earth PE Protective earth PE Protective earth bl blue N N sw b...

Страница 22: ...on site B2 CEE plug not assembled according to IEC standard 60309 2 F1 Mains fuse on site by customers F2 Mains fuse on site by customers F3 Mains fuse on site by customers S1 Main switch on site by c...

Страница 23: ...plate PD 4 D and PD 8 D Mains connection through on site switch S4 all pole separa tion from the mains possible 14 1 3 PD 4 D and PD 8 D for 220 V 3 at Mains supply 230 V 3 PE 50 60 Hz see nameplate P...

Страница 24: ...plate PD 4 D and PD 8 D Mains connection through on site switch S4 all pole sepa ration from the mains possible 14 1 5 PD 4 D and PD 8 D for 400 V 3 at Mains supply 400 V 3 N PE 50 60Hz see nameplate...

Страница 25: ...09 2020 Puridest Bi Distillation Apparatus 25 32 15 List of spare parts When ordering spare parts please always state type and serial number of the unit in question...

Страница 26: ...15101 Hose connector 0015115 Hose connector 10 A000115 Hose tap 11 A000084 Heating element Mono stage 2 0k W PD 2 D 1 pce PD 4 D 2 pcs and PD 8 D 3 pcs 12 0014516 Lock for housing 13 0012308 Connectio...

Страница 27: ...Apparatus without illustration Separate Water Supply with solenoid valve Only factory installation into models PD 2 D to PD 8 D is possible When the storage tank is full the solenoid valve switches o...

Страница 28: ...filter and fillings mounted completely Part No A000136 Part No A000135 Hose Set consisting of hoses for water inlet and outlet length 1 5 m including hose clips Hose Set Hose Set for seperate water s...

Страница 29: ...09 2020 Puridest Bi Distillation Apparatus 29 32...

Страница 30: ...30 32 Puridest Bi Distillation Apparatus 09 2020...

Страница 31: ...09 2020 Puridest Bi Distillation Apparatus 31 32...

Страница 32: ...LAUDA GFL Gesellschaft f r Labortechnik mbH Schulze Delitzsch Stra e 4 5 30938 Burgwedel Deutschland Tel 49 0 5139 9958 0 Fax 49 0 5139 9958 21 E Mail info lauda gfl de Internet www GFL de...

Отзывы: