Puridest Glass Distillation Apparatus
9 / 34
5
Operating Voltage
The Distillation Apparatus Models PD 2 G, PD 4 G and PD 2 DG must be connected to the mains supply through a cor-
rectly installed shock-proof socket. The models PD 8 G and PD 4DG must be connected to the mains supply through a
correctly installed on-site main switch, or, permanently connected, through a CEE plug in conformity with IEC standard
60309-2.
The unit is a protection class I electrical appliance, a connection to the protective earth is to be ensured.
For information on the value of the required mains fuse see Technical Data of these instructions.
The electrical connection must ensure an all-pole separation from the mains at any time. Shock-proof
sockets or main switches to separate the Distillation Apparatus from the mains must be installed in such a
way as to ensure clear identifiability and easy accessibility at any time.
The mains connection cable must not touch any hot surfaces of the unit. It may not lead underneath
the unit. The main switch of the Distillation Apparatus must be off (position O). The voltage on the
nameplate (at the left-hand side of the unit) must be identical to the mains voltage. If they are identical,
connect the unit to the mains.
Please also refer to chapter 13 of these operating instructions "Connection to the Mains"..
Содержание PD 2 G
Страница 1: ...Operating Instructions Puridest Glass Distillation Apparatus PD 2 G PD 4 G PD 8 G PD 2 DG PD 4 DG...
Страница 2: ...2 34 Puridest Glass Distillation Apparatus 09 2020...
Страница 4: ...4 34 Puridest Glass Distillation Apparatus 09 2020...
Страница 20: ...20 34 Puridest Glass Distillation Apparatus 09 2020 PD 2 G for 230 V 50 60 Hz PD 4 G for 230 V 50 60 Hz...
Страница 22: ...22 34 Puridest Glass Distillation Apparatus 09 2020 PD 2 DG for 230 V 50 60 Hz PD 4 DG for 220 V 3 PE 50 60 Hz...
Страница 32: ...32 34 Puridest Glass Distillation Apparatus 09 2020...
Страница 33: ...09 2020 Puridest Glass Distillation Apparatus 33 34...